SM150T User Manual 1.0
Troubleshooting
32
Water reading
Fully immerse the pins in water and measure the output in volts.
In the UK the sensor will typically read about 1.5 volts in tap water (because
the salinity is typically 50mS.m-1).
The “water reading” will depend on the salinity of your local water.
Note: HH150T
meter indicates “TOO WET” above 1.5V or 85% vol.
Graph: showing the effect of salinity on SM150T sensor output when
fully immersed in water with no soil present.
0.00%
10.00%
20.00%
30.00%
40.00%
50.00%
60.00%
70.00%
80.00%
90.00%
100.00%
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
0
100
200
300
400
500
600
O
u
tp
u
t
(%
vo
l)
O
u
tp
u
t
(V
o
lt
s)
Conductivity EC
p
(mS.m
-1
)
SM150T fully immersed in water - as read
by HH150 and HH2 meters
Volts
% vol (mineral)
%vol (organic)
"TOO WET"
Soil moisture
readings are not
correct when no
soil is present i.e. at
100% vol.
SM150T tables and
polynomial
constants are
optimised at 220
mS.m
-1
for soil
moisture values
below 70%vol