SM150T User Manual 1.0
Troubleshooting
31
Functional check
Air reading
Hold the SM150T in air and away from other objects and take a reading using
an HH150 or HH2 meter or voltmeter or a logger with no more than 5m of
cable.
Warning: Do not touch the pins
A typical electrostatic discharge from your body can create a
temporary -10mV offset in sensor readings lasting an hour.
In air an SM150T gives an output of 0 ±4mV.
Note: the HH150 reports under-range if the reading is less than
zero.
Mid range reading
– dip rod tips in water
If you wish to take a quick reading to check the sensor is
working you can dip the sensor into water.
With the pins half-immersed in tap water an HH2 set to read an
SM150T with soil type set to Organic should give a reading in the range 80 to
100%vol.