21
10.2.2 Temperature probes Pt100 using SICRAM module
Model
Type
Application range
Accuracy
TP472I
Immersion
-196°C…+500°C
±0.25°C (-196°C…+350°C)
±0.4°C (+350°C…+500°C)
TP472I.0
Immersion
-50°C…+400°C
±0.25°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP473P.0
Penetration
-50°C…+400°C
±0.25°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP474C.0
Contact
-50°C…+400°C
±0.3°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP475A.0
Air
-50°C…+250°C
±0.3°C (-50°C…+250°C)
TP472I.5
Immersion
-50°C…+400°C
±0.3°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
TP472I.10
Immersion
-50°C…+400°C
±0.3°C (-50°C…+350°C)
±0.4°C (+350°C…+400°C)
Common characteristics
Resolution
0.1°C
Temperature drift @20°C
0.003%/°C
10.2.3 - 4 wire Pt100 Probes
Model
Type
Application range
Accuracy
TP47.100
4 wire Pt100
-50…+400°C
Class A
Common characteristics
Resolution
0.1°C
Temperature drift @20°C
0.003%/°C