![Delta OHM HD2304.0 Manual Download Page 13](http://html1.mh-extra.com/html/delta-ohm/hd2304-0/hd2304-0_manual_2478328013.webp)
- -
13
4.2.3 Direct connection of 4 wire Pt100 sensors
4 wire Pt100
sensors
can
be soldered directly
to the pins of the flying female connector
without making use of the TP47 board. The 4
wires of the Pt100 sensors have to be soldered
as indicated in the figure on the left. In order
to use this type of probe it is necessary to set
up the menu item “Probe Type” as described
at page 9. The P100 probe is recognized upon
turning on the instrument: connect the probe
when the instrument is switched off and then
turn it on.
Sensor
Direct soldering to the connector
Pt100 4
wires
View of the soldering side
of the flying female connector