Delta OHM HD2304.0 Manual Download Page 15

-      - 

15 

6.

 

INSTRUMENT

 

SIGNALS

 

AND

 

FAULTS 

The  following  table  lists  all  error  indications  and  information  displayed  by  the  instrument  and 
supplied to the user in different operating situations:   
 

Display indications

 

Explanation

 

PROB 
COMM LOST

 

This  appears  if  the  SICRAM  module  connected  to  the  channel  has  already  been 
detected by the instrument, but is disconnected. At the same time an intermittent beep 
is issued. 

 

OVER

 

Measurement overflow: this appears if the pressure sensor exceeds the limit of 120% of 
the bottom scale nominal value. Over 125%, the display indicates ERR.

 

Measurement  overflow:  this  appears  if  the  external  temperature  probe  is  measuring  a 
value exceeding the set measuring range.

 

ERR

 

This appears in the menu if a module has already been detected by the instrument, but 
is disconnected.

 

PROB 
ERR

 

A probe with SICRAM module has been inserted when not admissible for that specific 
instrument.

 

SYS 
ERR 

Instrument  management  program  error.  Contact  the  instrument’s  supplier  and 
communicate the numeric code # reported by the display.

 

CAL 
LOST

 

Program error: it appears after turning on for a few seconds. Contact the instrument’s 
supplier.

 

BATT TOO LOW 
CHNG NOW

 

Indication of insufficient battery charge appearing on turning on. The instrument issues 
a long beep and turns off. Replace the batteries.

 

 

 

### BAR ABS

 

 probe ###bar absolute 

 

### BAR DIFF

 

 probe ###bar differential

 

### BAR GAUG

 

 probe ###bar relative 

 

### BAR SG

 

 probe ###bar relative compared to 1 bar

 

### mBAR ABS

 

 probe ###mbar absolute 

 

### mBAR DIFF

 

 probe ###mbar differential

 

### mBAR GAUG

 

 probe ###mbar relative 

 

PRBE_SER #### ####

 

 serial number #### #### of the probe connected to the input

 

ERR

 

 error

 

FUNC CLRD

 

 max, min and average values cleared

 

NEW_ PROB_DET

 

 new probe detected at the input

 

OVER

 

 maximum limit exceeded

 

PLS_EXIT  >>>  FUNC 
RES_FOR_FACT ONLY

 

please exit using ESC >>> function reserved to factory calibration

 

PRES_REL_TO_ZERO 
ENTR_TO_MENU

 

press REL to reset differential probe or ENTER to access menu

 

prob ERR

 

 error – unexpected probe

 

SYS ERR #

 

 program error number #

 

 

Summary of Contents for HD2304.0

Page 1: ...n this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual cannot be leg...

Page 2: ...2 Manometer Thermometer HD2304 0...

Page 3: ...tion in the menu increases the current value 8 UNIT MENU key it allows selection of the unit of measurement when pressed together with the DATA key it allows to open the menu 9 ZERO performs the zero...

Page 4: ...FAULTS 15 7 INSTRUMENT STORAGE 16 8 LOW BATTERY WARNING AND BATTERY REPLACEMENT 17 8 1 WARNING ABOUT BATTERY USE 17 9 NOTES ABOUT WORKING AND OPERATIVE SAFETY 18 10 TECHNICAL CHARACTERISTICS 19 10 1...

Page 5: ...bes The probes are fitted with the SICRAM automatic detection module with the factory calibration settings already being memorized inside The units of measurement for the measurable quantities are Pa...

Page 6: ...on can be disabled by simultaneously pressing this key and the HOLD key when turning the instrument on During turning on should no probes be connected the NO_PRBE_SER_NUM message is displayed in the l...

Page 7: ...trument is off the previously memorized data are cleared Upon turning on the instrument automatically starts memorizing the MAX MIN and AVG values To reset the previous values and start with a new mea...

Page 8: ...nnected to the instrument s input see chapter 3 the menu is opened by pressing simultaneously DATA UNIT the first item of the instrument programming menu will appear use the and arrows respectively lo...

Page 9: ...te are displayed alternately in the secondary line temperature probes Pt100 complete with SICRAM module the Auto message is displayed in the central part of the display direct 4 wire Pt100 temperature...

Page 10: ...nts The threaded connection is protected by a plastic cap Put it back after use as it protects the pressure cell from foreign bodies IMPORTANT Ensure the probe s bottom scale is higher than the pressu...

Page 11: ...probe NOTE when measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe In order to perform a correct contact measuremen...

Page 12: ...00 3 wires Pt1000 Ni1000 Not Used Caution Before welding pass the probe cable through the fairlead and gasket 6 Weld the wires as shown in the table Sensor Board connection Jumper to close Pt100 4 wir...

Page 13: ...ve to be soldered as indicated in the figure on the left In order to use this type of probe it is necessary to set up the menu item Probe Type as described at page 9 The P100 probe is recognized upon...

Page 14: ...erform It has high levels of uncertainty and depends on the ability of the operator 9 Temperature measurements on non metal surfaces usually require a great deal of time due to the low heat conductivi...

Page 15: ...ecific instrument SYS ERR Instrument management program error Contact the instrument s supplier and communicate the numeric code reported by the display CAL LOST Program error it appears after turning...

Page 16: ...n places where humidity is high the instrument may be exposed to direct sunlight the instrument may be exposed to a source of high temperature the instrument may be exposed to strong vibrations the in...

Page 17: ...strument back on To replace the batteries proceed as follows 1 switch the instrument off 2 unscrew the battery cover counter clockwise 3 replace the batteries 3 1 5V alkaline batteries type AA 4 screw...

Page 18: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Page 19: ...y 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20 A Connections Input for probes 8 pole male DIN45326 connector Unit of Measurement Pa hPa kPa mbar bar atm mmHg mmH2O k...

Page 20: ...80 C BSP 400 mbar 800 mbar 900 mbar 0 1 mbar TP704 400MBGI 0 25 FSO 10 80 C BSP 500 mbar 1500 mbar 1800 mbar 0 1 mbar TP705 500MBD 0 25 FSO 0 60 C Tube 5mm 600 mbar 1200 mbar 1500 mbar 0 1 mbar TP704...

Page 21: ...400 C TP474C 0 Contact 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP475A 0 Air 50 C 250 C 0 3 C 50 C 250 C TP472I 5 Immersion 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP472I 10 Immersion 50 C...

Page 22: ...TP472I 0 Pt100 sensor immersion probe Stem 3 mm length 230 mm Cable length 2 metres TP473P 0 Pt100 sensor penetration probe Stem 4 mm length 150 mm Cable length 2 metres TP474C 0 Pt100 sensor contact...

Page 23: ...i campioni internazionali e nazionali delle unit del SIT garantita da una catena di riferibilit ininterrotta che ha origine dalla taratura dei campioni di laboratorio presso l Istituto Primario Nazion...

Page 24: ...ectronic devices with the following symbol cannot be disposed in the public dumps According to the Directive UE 2002 96 EC the European users of electric and electronic devices are allowed to give bac...

Reviews: