Delta OHM HD2304.0 Manual Download Page 10

-      - 

10 

4.

 

PROBES

 

AND

 

MEASUREMENTS 

The  instrument  measures  absolute,  relative  and  differential  pressure  using  the  PP471  SICRAM 
module  and  the  TP704  and  TP705  series  probes.  It  also  measures  the  temperature  with  a  Pt100 
sensor fitted with SICRAM module, or with direct 4 wire Pt100 probes.  

The  SICRAM  module  acts  as  an  interface  between  the  sensor  on  the  probe  and  the  instrument. 
There  is  a  microprocessor  circuit  with  a  permanent  memory  inside  the  module  that  enables  the 
instrument to recognize the type of probe connected and to read its calibration information.  

The probes are detected during turn on, and this cannot be performed when the instrument is 
already on, therefore if a probe is connected and the instrument is on, it is necessary to turn it 
off and on

The probes fitted with SICRAM module are calibrated in the factory; no calibration is required by 
the user. 
 

4.1

 

PP471

 MODULE FOR PRESSURE MEASUREMENT

 

The PP471 module acts as an interface between the TP704 and TP705 series Delta Ohm probes and 
the instrument.  
The 

TP704

 series pressure probes have a 1/4" BSP threaded male connection, and must be screwed 

into the system using the necessary sealing gaskets. 
The 

TP705

 series have two Ø5 connections where the suitable tubes are inserted in order to perform 

the desired measurement.  

Caution! 

Please pay careful attention to the joint pressure sealing; use suitable gaskets and joints. 

The threaded connection is protected by a plastic cap. Put it back after use as it protects the pressure 
cell from foreign bodies. 

IMPORTANT!  Ensure  the  probe's  bottom  scale  is  higher  than  the  pressure  that  will  be 
measured. In case this value is unknown, start by using higher-capacity probes.

 

For  every  pressure  probe  a  range  of  overload  pressure  and  a  burst  pressure  are  declared: 
pressures  in  the  range  of  overload  pressure  don’t  cause  the  break  of  the  sensor  but  the 
declared  accuracy  can  be  overcome.  Pressure  over  the  stated  burst  pressure  may  break  the 
sensor. The application of higher pressures than the overload pressure limit, although lower 
than the burst pressure, may produce permanent damage to the probe (e.g. offset shift). Never 
exceed the stated burst pressure. 

Please see the technical characteristics of the probes on paragraph  “Technical characteristics”.  

Upon  turning  on  the  instrument  automatically  detects  the  PP471  module.  The  probe's  type 
(absolute, relative or differential) and bottom scale value are detected even when the instrument is 
on (see chapter 3).  

To change the instantaneous or peak value unit of measurement, press UNIT. The following units of 
measurement are available:  

Pa, hPa, kPa, mbar, bar, atm, mmHg, mmH

2

O, kgf/cm

2

, PSI, inchHg. 

Some units of measurement require a degree of multiplication:  

 

 the "-3" symbol indicates the displayed value must be divided by 1,000; 

 

 the "3" and "6" symbols  indicate the displayed value must be multiplied respectively by 1,000 
or by 1,000,000. 

The instrument does not automatically detect the probes which are not fitted with SICRAM module. 

 

Summary of Contents for HD2304.0

Page 1: ...n this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual cannot be leg...

Page 2: ...2 Manometer Thermometer HD2304 0...

Page 3: ...tion in the menu increases the current value 8 UNIT MENU key it allows selection of the unit of measurement when pressed together with the DATA key it allows to open the menu 9 ZERO performs the zero...

Page 4: ...FAULTS 15 7 INSTRUMENT STORAGE 16 8 LOW BATTERY WARNING AND BATTERY REPLACEMENT 17 8 1 WARNING ABOUT BATTERY USE 17 9 NOTES ABOUT WORKING AND OPERATIVE SAFETY 18 10 TECHNICAL CHARACTERISTICS 19 10 1...

Page 5: ...bes The probes are fitted with the SICRAM automatic detection module with the factory calibration settings already being memorized inside The units of measurement for the measurable quantities are Pa...

Page 6: ...on can be disabled by simultaneously pressing this key and the HOLD key when turning the instrument on During turning on should no probes be connected the NO_PRBE_SER_NUM message is displayed in the l...

Page 7: ...trument is off the previously memorized data are cleared Upon turning on the instrument automatically starts memorizing the MAX MIN and AVG values To reset the previous values and start with a new mea...

Page 8: ...nnected to the instrument s input see chapter 3 the menu is opened by pressing simultaneously DATA UNIT the first item of the instrument programming menu will appear use the and arrows respectively lo...

Page 9: ...te are displayed alternately in the secondary line temperature probes Pt100 complete with SICRAM module the Auto message is displayed in the central part of the display direct 4 wire Pt100 temperature...

Page 10: ...nts The threaded connection is protected by a plastic cap Put it back after use as it protects the pressure cell from foreign bodies IMPORTANT Ensure the probe s bottom scale is higher than the pressu...

Page 11: ...probe NOTE when measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe In order to perform a correct contact measuremen...

Page 12: ...00 3 wires Pt1000 Ni1000 Not Used Caution Before welding pass the probe cable through the fairlead and gasket 6 Weld the wires as shown in the table Sensor Board connection Jumper to close Pt100 4 wir...

Page 13: ...ve to be soldered as indicated in the figure on the left In order to use this type of probe it is necessary to set up the menu item Probe Type as described at page 9 The P100 probe is recognized upon...

Page 14: ...erform It has high levels of uncertainty and depends on the ability of the operator 9 Temperature measurements on non metal surfaces usually require a great deal of time due to the low heat conductivi...

Page 15: ...ecific instrument SYS ERR Instrument management program error Contact the instrument s supplier and communicate the numeric code reported by the display CAL LOST Program error it appears after turning...

Page 16: ...n places where humidity is high the instrument may be exposed to direct sunlight the instrument may be exposed to a source of high temperature the instrument may be exposed to strong vibrations the in...

Page 17: ...strument back on To replace the batteries proceed as follows 1 switch the instrument off 2 unscrew the battery cover counter clockwise 3 replace the batteries 3 1 5V alkaline batteries type AA 4 screw...

Page 18: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Page 19: ...y 200 hours with 1800mAh alkaline batteries Power absorbed with instrument off 20 A Connections Input for probes 8 pole male DIN45326 connector Unit of Measurement Pa hPa kPa mbar bar atm mmHg mmH2O k...

Page 20: ...80 C BSP 400 mbar 800 mbar 900 mbar 0 1 mbar TP704 400MBGI 0 25 FSO 10 80 C BSP 500 mbar 1500 mbar 1800 mbar 0 1 mbar TP705 500MBD 0 25 FSO 0 60 C Tube 5mm 600 mbar 1200 mbar 1500 mbar 0 1 mbar TP704...

Page 21: ...400 C TP474C 0 Contact 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP475A 0 Air 50 C 250 C 0 3 C 50 C 250 C TP472I 5 Immersion 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP472I 10 Immersion 50 C...

Page 22: ...TP472I 0 Pt100 sensor immersion probe Stem 3 mm length 230 mm Cable length 2 metres TP473P 0 Pt100 sensor penetration probe Stem 4 mm length 150 mm Cable length 2 metres TP474C 0 Pt100 sensor contact...

Page 23: ...i campioni internazionali e nazionali delle unit del SIT garantita da una catena di riferibilit ininterrotta che ha origine dalla taratura dei campioni di laboratorio presso l Istituto Primario Nazion...

Page 24: ...ectronic devices with the following symbol cannot be disposed in the public dumps According to the Directive UE 2002 96 EC the European users of electric and electronic devices are allowed to give bac...

Reviews: