23
interval and then confirm it with the <ENTER> key. Press <ESC> > to quit without making
any changes.
1
<ESC> exit/cancel
current print interval
This parameter influences the functions of the immediate printing of data:
Screen
,
RCD+
and
Rawdata
(
see the functions “Screen”, “RCD+” and “ALL” from p.90
)
5)
C
ALIBRATE
Manages the calibration functions of the probes connected to the instrument. The
probe+instrument calibration function may be protected by a password (
see the function on page
15 “1-1) Reserved function lock”
). If user level is enabled, when you enter the function with
the <Enter> key a list of the probes connected to the instrument inputs appears: each probe is
identified by its serial number and by the type of measurement. Selecting one by pressing the
number next to each probe takes you to the calibration menu.
5
1
<ESC> exit/cancel
calibration of the
probe connected
to input A
is started
0
1
2
3
4
5
7
9
6
8
Each type of probe has its own calibration procedure:
see the description of the various probes
and the respective calibration operations from p.28
.
Instead, if user level is disabled, it will not be possible to calibrate any probe; in this case the fol-
lowing message will appear: “Sorry! This operation is RESERVED”. Press <ESC/CLR> to quit.
To proceed with calibration you must enable user level by typing in the user password and then
repeat the operation.
5
<ESC> exit/cancel
0
1
2
3
4
5
7
9
6
8
6)
R
ESET
With this command the instrument parameters are returned to default conditions. The variables
that are reset are the date, time, the configurable options protected by password, the baud rate for
serial communication, the print options and the logging functions.