33
<ESC> exit/cancel
At this point it is possible to insert the type of thermocouple by pressing the corresponding nu-
merical key: for example, if you want to set thermocouple type J, press the numerical key
<2/LOG> and confirm with the <ENTER> key when the next screen appears.
2
9=x
(reset type)
(
works as K)
x
now at J
Offset Compensation
Press <ESC/CLR> to skip this step. Bring the calibration bath to the temperature used for com-
pensating offset at 0°C and immerse the probe in it (or the two probes of the double module).
Wait until the probes have reached the bath temperature and then press <ENTER>. The follow-
ing screen appears:
Zero Point
0.0°c
0.55°c
0.30°c
Wait until the temperatures indicated for channels A and B have stabilised. Using the
Up
and
Down
arrows correct the calibration value proposed by the instrument (Target) and make it coin-
cide with the bath temperature measured by the reference thermometer. To confirm, press
<ENTER>: in this way the temperature values of the thermocouple probe (or of the two probes if
the double module is connected) automatically go to the value indicated as the “Target” and