81
The functions Logging (key <2/LOG> and Record (key <6/RCD>)
are active at the same time. To end the Logging function press the
key <2/LOG>. To end the Record function, press the key <6/RCD>
and then the function key <F2>–
rcdSTOP
–.
(See the “Logging”
function on p.88 and the function Record on p.86).
The Logging function (key <2/LOG>) and the immediate printing
sub-function RCD+ (key <4/SerialOut>
Æ
key <F2>) are active at
the same time. To end the Logging function press the key
<2/LOG>. To end the sub-function RCD+, press the key
<4/SerialOut > and then the function key <F1> – STOP printing –.
(See the “Logging” function on p.88 and the RCD+ function on
p.91).
The Logging function (key <2/LOG>) and one of the two sub-
functions for immediate printing,
Screen
(key <4/SERIALOUT>
Æ
key <F1>) or
Rawdata
(key <4/SERIALOUT>
Æ
key <F1>),
are active at the same time. To end the Logging function press the
key <2/LOG>. To end the sub-functions
Screen
or
Rawdata,
press
the key <4/SERIALOUT> and then the function key <F1>–
STOP printing
–.
(See the Logging function on p.88, the Screen function on p.90 and
the Rawdata function on p.92).
Automatic starting of the Logging function has been set. (
See the
function “Start/Stop time” on p.18
)
r
With the SICRAM pressure module PP471, indicates a relative
measure (See the chapter “PP471 ELECTRONIC MODULE FOR
PRESSURE MEASUREMENT”)
WARNING! CHANGE
OF PROBES
DETECTED – Press
NOW any key to choose
settings or wait to self-
config
This message appears when switching on the machine or leaving
the menu, when a variation is detected at the inputs because a probe
has been connected or removed.
(See on p.28 the introduction to
the chapter on probes)
.
COM FAILURE
Communication failure.
This message appears when, once the in-
strument is already on, a module is disconnected: the instrument
consequently indicates that there is no communication with the in-
volved channel.
LOW BATTERY
Log refused
Logging cannot be starter as the battery level is too low.