Appendix I
Mains Isolation
The Trigno Base Station is provided with Medical Grade isolated
power supply which is compliant with IEC60601 series of
harmonized standards for Medical Devices. However, full
compliance with IEC60601-1 Basic Safety for Medical Devices
requires that the PC operating the software be isolated as well. This
stems from the basic requirement to have all patients electrically
isolated from equipment within their reach, and since the PC
running the Trigno Software is conceivably within their reach, it
too must be isolated.
Delsys does not supply isolation transformers for Personal
Computers and their peripherals.
Delsys recommends model IS1000HG manufactured by Tripp
Lite (www.tripplite.com) for this task. This device is a medial
grade isolation transformer capable of delivering up to 1000 W.
A smaller similar version for 500W is also available (IS500HG).
Similar products compliant with IEC60601-1 are acceptable.