Page: 60 of 102
16.
Fault finding
See Section 2.
The OMD-2008 Series will indicate several malfunctions in the status line of the display.
Pressing the
“OK”
button will display additional information.
Status
OK
Reading
0..15
15..49
EE
System Alarm LED
Green/ Blinking
Alarm Circuit 1,2
No Alarm
Alarm
Reason
Low oil content.
Normal Operation.
Sample reading is
out of range:
-oil content too
high.
-dirty Sample
Glass Tube.
-Air bubbles in
sample stream.
Recommendations
-Continue operation.
-Apply standard
maintenance
schedule.
Resume normal
maintenance
schedule.
-Wait until oil
content is within
the range.
-Clean Sample
Glass Tube.
Status
FW!
Flow!
E-VALVE?
Reading
0..49/ EE
Any
System Alarm LED
Green/ Blinking
Red
Alarm Circuit 1,2
Alarm
Reason
- Freshwater is
enabled.
- Remote fresh
water input is linked
19 -> 20).
-Terminals 15&16
open.
- A magnet is
missing.
- Flow sensor
unsatisfied.
- Magnet sensor
defective.
EV FW Valve
position not
detected.
(For EV Units only)
Recommendations
- Switch back to
sample stream.
Check sample flow
rate.
Check connectors
and wiring.
Summary of Contents for OMD-2008
Page 28: ...Page 27 of 102 5 5 Cabling OMD 2008 EV Figure 9 Bottom View OMD 2008 EV ...
Page 84: ...Page 83 of 102 ...
Page 85: ...Page 84 of 102 ...
Page 86: ...Page 85 of 102 ...
Page 87: ...Page 86 of 102 ...
Page 88: ...Page 87 of 102 ...
Page 89: ...Page 88 of 102 ...
Page 90: ...Page 89 of 102 ...
Page 91: ...Page 90 of 102 ...
Page 92: ...Page 91 of 102 ...
Page 93: ...Page 92 of 102 ...
Page 94: ...Page 93 of 102 ...
Page 95: ...Page 94 of 102 ...