Installation OMD-2008 EVA
The OMD-2008 EVA is fitted with the Auto Clean System. Installation of the OMD-2008
EVA can be carried out in a few steps. Please refer to the following drawing for mounting
dimensions.
Install the Air Regulation Filter Unit close enough to the OMD-2008 EVA for connection of
the air pipe.
a)
Connect an air supply of 4 to 6 bar to Air Regulation Filter Unit.
b)
Set the Air Regulation Unit output to a pressure of approximately 4 bar typical. Too
high-pressure settings can cause unnecessary wear. Low pressure may make
operation unreliable.
c)
Insert the wiper into the cleaner. Check the Wiper Seal for damage.
d)
Put Automatic Cell Cleaning Device onto Measuring Cell. Make sure it is completely
screwed in. If it is only partially screwed in, flow rate regulation will not work and may
cause unstable operating conditions for the instrument.
e)
Connect air hose to Automatic Cell Cleaning Device.
Figure 11. Installation OMD-2008 EVA