DDS CALORIMETERS OPERATIONS MANUAL V1.1
2022
Website:
www.ddscalorimeters.com
39
Digital Data Systems (Pty) Ltd.
CAL3K Advanced Bomb Calorimeters
www.ddscalorimeters.com
Enter any output form with the letters A-H and see the printing of the header and today’s data.
Or print result from memory (any day) with:
69
PRT RESLTS FORM: 0
Results are printed. If no results are in memory, then put one in:
77
SIMULATE RESULT 0
20.3
PRINTING A FORM HEADER
A header is a text string, which identifies each data column. The identifications are ‘cryptic’ but you get
the idea:
56
PRT A-H HEADER
0
20.4
PRINT FILTER
If required a filter can be applied before printing. Results (memory entries), which are:
Calibrations
Faulty determinations
can be suppressed with the following command:
55
SET FILTER C,F:
1
20.5
PRINTING RESULTS FROM MEMORY
All results in memory can be printed with the provision:
C
Calibrations
R
Results
YES, Calibrations and Results
CR
Calibrations and results
62
SET PRT MEMO.CR: 1
20.6
PRINTING RESULTS BY GID
If required all results with the specified GID can be printed.
64
PRT RESULT GID
0
Note
that the FULL GID is required. If GID = Date is set then all results of a particular date are printed.
The print format may not show the GID or not all of it.