- 14 -
Shutter Error: Please,
restart the laser.
Improper operation of the
laser resonator shutter.
Restart the laser. If the
problem persists, turn the
system OFF and call Customer
Service for assistance.
Error: No water flow.
Please, call for service!
There is insufficient water
flow in the cooling system.
Check water level in the
cooling tank. Add distilled
water, if necessary. Restart the
laser. If the problem persists,
turn the system OFF and call
Customer Service for
assistance.
Emergency Button
depressed. Please release the
button and restart the laser.
Laser was turned off using the
Emergency Button.
Release (rotate CW and pull
up) the Emergency Button and
restart the laser.
Charger Fault: Please, call
for service!
Capacitor charger can not
charge to required voltage or
charging is too slow.
Restart the laser. If the
problem persists, turn the
system OFF and call Customer
Service for assistance.
OverFluence Error. Please,
restart the laser.
The measured Output Fluence
was more than 20% higher
than pre-set by the operator.
Restart the laser. If the
problem persists, call
Customer Service for
assistance.
Overcharge Error: Please,
call for service!
The measured charge on the
capacitor bank was more than
10% higher than ordered by
controller.
Critical system fault. Please,
call Customer Service for
assistance.
Charger Overheat Error:
Please, call for service
!
Capacitor charger temperature
is too high.
Make sure that ambient room
temperature is below 25
°
C
. Do
not turn the laser off.
Please,
allow time for the system to
cool down and restart the
laser. If the problem persists,
call Customer Service for
assistance.
Main Switch Error: Please,
call for service!
One of the communication
relays was not operating
properly.
Please, restart the laser. If the
problem persists, call
Customer Service for
assistance.
System overheat, T=xx.x.
Please, call for service.
Water temperature in the
cooling system was above
critical level.
Please, allow time for the
system to cool down and
restart the laser. Make sure
that cooling system fan has
turned on. If not, please, call
Customer Service for
assistance.