- 4 -
Laser Handpiece
The laser handpiece is equipped with the distance gauge, which provides fixed laser beam
spot size on the treatment area. The spot size and the fluence level can be changed from
the laser control panel. The motorized zoom lens position is electronically recognized by
the system and is reflected as the Laser Beam Spot Size (
Beam Size
) on the LCD display.
The system processor will calculate the energy fluence delivered to the patient skin in
accordance with selected laser beam spot size.
Component
Description
Handpiece Handle
The user holds the laser handpiece by the
body-handle during treatment. The optional
removable plastic handle with pulse trigger
switch is also available.
Protective Window (Output
Aperture)
The handpiece output window serves as the
laser aperture. It protects the main focusing
lens from contamination during treatment.
The collimating lens inside the handpiece
changes the laser beam divergence out of the
fiber, and in combination with imaging zoom
lens provides fixed laser beam spot size at
the certain distance from the handpiece.
Fiber Cable Umbilical Jacket
Contains optical fiber, electrical cables and
cooling air tubing. The reinforced umbilical
jacket protects optical fiber from mechanical
damage. It also prevents the fiber from being
bent to too small radius during treatment.
Warning! The umbilical jacket can
protect the optical fiber to the certain
extent only. Do not apply force to try to
bend the cable to the smaller radius, than
the jacket easily allows or by pulling on
the fiber by the handpiee. This can lead to
the permanent fiber cable damage, which
is not covered by Limited Warranty.
Distance Gauge
Provides fixed distance to get correct beam
spot size on the treatment area. Second
gauge also serves as a guide and output
nozzle for the cooling air jet.