C1000 Series Service & Maintenance Manual
Theory of Operation
Theory of Operation
THEORY OF OPERATION
While Running
To start the compressor, the user shall activate the compressor head or heads with the power switches located on the front center
of the machine. The selected motors will then begin to run and build pressure in the tank and piping system.
A. The air is taken in through the intake
fi
lters, located on the rear of each compressor cylinder. The dirt and dust is
fi
ltered out
of the intake
fi
lter to 5 microns in this process.
B. The air is then compressed with the pistons in each cylinder, and then travels through the after-cooler (copper) tubing, which
cools the air to help condense any moisture “humidity” taken in during the process of compression.
C. The wet / dirty compressed air and condensed moisture then travel through a coalescing
fi
lter assembly. This
fi
lter allows the
condensed moisture to fall to the bottom of the
fi
lter bowl via gravity, where it will be automatically expelled via the
fl
oat
drain contained in the bottom of the
fi
lter bowl. The compressed air also travels through the coalescing
fi
lter element, which
removes 99.998% of all particles down to 0.01 microns. A
fi
lter change indicator located on the top of the
fi
lter housing
shows the condition of the element, and alerts the operator as to when the element is in need of replacement. GREEN
indicates the element has more useful life, while RED indicates the element is in need of replacement.
NOTE:
The indicator only functions while compressed air is traveling through the
fi
lter with compressor running. The indicator
will ALWAYS read “GREEN” with compressors not running.
D. The compressed air then travels through the bottom of the desiccant “drying chamber”, where the remaining contained
moisture is reduced to a pressure dewpoint of -100
°
F, and then travels into the main storage tank for use in the dental
operatory.
E. When reaching its cut-off pressure of 100psi, the pressure switch will open and the compressor motors will stop.
NOTE
: The time operated dryer purge valve DOES NOT open and purge when compressor stops, as it does on most competitive
brands. See Theory of Operation “While Not Running” for more information as to the proper operation of our purge valve.
SECTION II
1.800.624.2793
5