3KP78 - 1.1
3000PLUS
WITH
5D78 I
NSTRUCTION
M
ANUAL V
. SB.3
T
HE
3000PLUS P
ANEL
M
ETER
Incorporating Daytronic’s
5D Series Signal Condition-
er Modules
, the 3000PLUS Panel Meter is a field-scal-
able indicator featuring operator-programmable signal
processing and PC/PLC communications. With a
durable front panel and secure screw terminals for all
power and I/O connections, this mechanically and elec-
trically rugged instrument is ideal for pump, motor,
hydraulic, and other high-noise monitoring applications.
Accepting the fully conditioned output of any standard
plug-in 5D module*, the 3000PLUS maintains signal
integrity to deliver accurately scaled analog output,
while sampling all data at 16-bit resolution. The data dis-
play provides selectable digital filtering for even greater
readout stability.
The operator can easily select any of three separate
output channels for display, as explained in Section 1.D:
•
the meter’s standard ±5-VDC scaled output
(
Channel 1
), representing measured engineering
units (after calibration)
•
the “auxiliary” DAC output (
Channel 2
), which
serves to monitor and process Channel 1, and to
generate Channel 3
•
the “scaled voltage” output (
Channel 3
), which is
continuously proportional to the reading of the auxil-
iary output (Channel 2). Channel 3 may be set to a
full scale of either ±5 or ±10 VDC, and is available
from the rear of the 3000PLUS as both voltage and
4-20 mA output (see Section 2.D).**
The auxiliary analog output (Channel 2) is used for
•
HI/OK/LO
limit monitoring
with selectable hystere-
sis windows, front-panel annunciation, and relay out-
puts for local process control (see Section 5.E for
full details)
•
high-speed positive or negative
peak capture
with
TTL-level “have peak” output, selectable “backout”
and “peak defeat” thresholds, and user-adjustable
leak rate (Section 5.B)
•
a
signal hold
to allow captured peaks and other
values to be transferred to computer for processing
(Section 5.C)
•
automatic application of a desired
tare offset
(Sec-
tion 5.D)
Separate logic inputs provide external control of peak
capture, signal hold, tare application, and the release of
latched limit violations.
As explained in Sections 3 and 4 of this manual, you
can quickly set up the 3000PLUS either via the simple
front-panel button menu or via the configuration soft-
ware supplied with the unit. Operator-entered ranges,
filters, calibration points, and other setup parameters
are always specific to the installed 5D Series condition-
er.
A standard RS232 interface operating at a fixed rate of
19.2K baud allows connection of an external PC for
instrument configuration and on-line monitoring. It may
also be used for direct communication of simple ASCII
mnemonic commands both for run-time reconfiguration
and for interrogation of current data and setup values.
Employing the run-time version of Microsoft
®
Access
2000, the
3KP CONFIGURATOR
software supplied
with the Model 3000PLUS makes short work of meter
setup. Communicating via the RS232 link, the Configu-
rator lets you define, store, edit, download, upload, and
manage any number of “configurations” for your
3000PLUS. As explained in Section 4, “off-line” configu-
rations can be easily created and downloaded, or you
can use the “Live Output Window” to view and modify
the present configuration of the connected 3000PLUS
instrument on a purely run-time basis.
The Configurator also lets you perform selected run-
time operations, including
•
both “absolute” and two-point (deadweight) calibra-
tion
•
viewing any of the meter’s three “live” analog out-
puts and adjusting it as desired
•
applying a signal hold command
•
releasing any and all latched limits
•
sending standard mnemonic commands to the
meter
•
activating the shunt resistor for verification of sensor
calibration data.
T
HE
M
ODEL
5D78 AC S
TRAIN
G
AGE
C
ONDITIONER
M
ODULE
Delivering a high-level filtered analog output of ±5 VDC,
the Model 5D78 is a single-channel strain gage condi-
tioner of phase-sensitive carrier-amplifier design.
While intended primarily for applications involving trans-
former-coupling to the transducer bridge (as with con-
ventional
rotary-transformer torque sensors
), the
Model 5D78 can also be used when high sensitivity is
required or where the electrical environment is espe-
1. I
NTRODUCTION
1.
A
G
ENERAL
D
ESCRIPTION AND
S
PECIFICATIONS
* “V” and “S” models may not be used with the 3000PLUS.
** The ±5-VDC output of the installed 5D78 conditioner (prior to
A/D conversion) is also available from the rear of the meter.