3KP78 - 2.1

3000PLUS 

WITH 

5D78 I

NSTRUCTION 

M

ANUAL V

. SB.3

2.  C

ONNECTIONS

The 3000PLUS requires a user-supplied external
source of 24 VDC, regulated to ±10%.  Nominal con-
sumption is 300 mA; maximum is 350 mA.*  The figure
below shows how the positive and negative power
leads are tied, respectively to the rear-panel 

+24 VDC

and 

POWER COMMON

terminals.  Local grounding is

not required.

NOTE: On every normal powerup, the 3000PLUS will
display (for two or three seconds) its 

firmware version

number

(“V x.x”) alternating with

2.

A

P

OWER

C

ONNECTIONS

A

 

/

 

D

COM

OUT

COM

A

 

/

 

D

INP

V

OUT

4-20

OUT

LOG

COM

HLD

PEK

TAR

LOG

COM

RCV

XMT

PWR

COM

+24

VDC

HI

HI

LO

LO

OK

OK

HAV

PEK

RLS

LAT

24 VDC

POWER

SUPPLY

Fig. 6

3000PLUS Power 
Connections

PLEASE NOTE: Some of the wiring diagrams in this section show only the male connector head-
ers on the rear of the 3000PLUS instrument (also shown in Fig. 3).  Each connected wire or jumper
is to be firmly secured to the corresponding 

SCREW TERMINAL

of the terminal block (supplied

with the meter) that plugs into the appropriate header.

C

ABLE

S

HIELDING

Proper shielding of cable wires or twisted pairs—as shown in Figs. 7 through 10—is strongly rec-
ommended to minimize the production of unwanted electrical noise from capacitive and inductive
effects.  

In the I/O cabling diagrams below, only the “connector end” of each cable shield is shown, as rep-
resented by a gray circle surrounding either a single wire or a TWISTED PAIR of wires within the
cable.  Unless otherwise stated, every shield should be grounded to the appropriate common or
ground terminal only at the connector end
.  The drain wire tying the connector end of the shield to
common/ground should be as short as possible. 

* An optional 18-W in-line power supply for the 3000PLUS (the

Model 3KPS1

) is available from Daytronic.  Contact the factory

for more information.

Summary of Contents for 3000PLUS

Page 1: ...d reproduced or used in any form or by any electronic mechanical or other means including photocopying and recording or in any information storage and retrieval system without permission in writing fr...

Page 2: ...has been properly installed If your 3000PLUS is equipped with a 5D Series conditioner model other than the 5D78 the corresponding instruction manu al should be consulted ALWAYS TURN OFF THE 3000PLUS B...

Page 3: ...ront Panel Setup Procedure 3KP78 3 4 4 SOFTWARE CONFIGURATION AND CALIBRATION 4 A Using the 3000Plus Configurator Software 3KP78 4 1 4 B Summary of Configurator Menus 3KP78 4 3 4 C Overview of OFF LIN...

Page 4: ...ain Gage Cabling under 20 ft in length 3KP78 2 3 Fig 8 b 8 Wire Strain Gage Cabling 20 ft or longer 3KP78 2 3 Fig 9 Analog Output Connections 3KP78 2 4 Fig 10 3000PLUS Logic I O Connections Fig 10 a L...

Page 5: ...plained in Sections 3 and 4 of this manual you can quickly set up the 3000PLUS either via the simple front panel button menu or via the configuration soft ware supplied with the unit Operator entered...

Page 6: ...nd power supply communi cations terminals 1 INTRODUCTION High ESD immunity and extensive EMI protec tion further assure data integrity in harsh industrial environments Because of normal cable loading...

Page 7: ...inal 0 5 V where 5 V Logic 1 AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA A A A A 1 70 4 41 7 17 0 22 Fig 1 3000PLUS Dimensions true 25 V without damage noise im...

Page 8: ...d excitation frequency for an excitation of 10 00 kHz it is 2000 Hz for 5 00 kHz it is 1400 Hz and for 3 27 kHz it is 1100 Hz See Table 2 in Appendix B for the practical ranges that apply to the 5D78...

Page 9: ...ction 1 D Setup Indicators see Section 3 B Channel and Limit Value Indicators see Section 1 D Communi cations Indicator see Section 2 B Fig 2 3000PLUS Front Panel Elements Limit Logic Relay Outputs se...

Page 10: ...ser s Panel with Cutout and Mounting Holes see Fig 5 A B 4 060 3 700 0 180 0 200 0 156 dia Edge of 3000PLUS instrument 1 45 0 775 Fig 4 3000PLUS Panel Mounting Fig 5 Panel Cutout Dimensions 3 Remove t...

Page 11: ...setup or run time basis as explained in Section 4 The DECIMAL POINT RESOLUTION of the data display for Channels 1 and 2 will always match that of the last entered FSU value 2 You can also specify a DI...

Page 12: ...tification 2 RG RANGE input range and scaling information including for the 5D78 module excitation frequen cy full scale mV V range desired full scale reading in units and desired decimal point resolu...

Page 13: ...perator is called on to select one of a set of discrete values for a given setup parameter this button is used to cycle through the display of those values for details see Entering a Setup Parameter i...

Page 14: ...clears any currently captured peak value by momentarily disabling and re enabling peak capture see Section 5 B SETUP FUNCTIONS Used during setup to 1 Acknowledge each setup parameter when it appears...

Page 15: ...ut does require full installation of Microsoft Access 2000 Runtime which is supplied with the Configurator and which takes approximately 32 MB of hard drive space If Microsoft Office 2000 or higher is...

Page 16: ...Microsoft Office Office so that it reads C Program Files Microsoft Office Office11 f Click OK g Now you can run the Configurator by clicking on 3000PLUS Configurator in the Programs list THE CONFIGURA...

Page 17: ...want to completely remove the 3KP Configurator and all its components answer Yes to uninstall or No to abort You may be asked whether you re sure you want to remove MSCOMM32 OCX and COMDLG32 OCX which...

Page 18: ...r of the 3000PLUS instrument also shown in Fig 3 Each connected wire or jumper is to be firmly secured to the corresponding SCREW TERMINAL of the terminal block supplied with the meter that plugs into...

Page 19: ...nications between the 3000PLUS and an external PC While 3000PLUS PC serial com munications will usually take place through the 3KP Configurator software described in Section 4 a termi nal emulation pr...

Page 20: ...er 20 ft in length CAL SENSE SIGNAL CAL ENABLE EXCITATION SENSE SENSE EXCITATION SHIELD 10 9 8 7 6 5 4 3 2 1 SIGNAL EXTRA WIRE paired with CAL SENSE unconnected at Transducer Connector NOT ISOLATED co...

Reviews: