background image

3KP78 - 1.7

3000PLUS 

WITH 

5D78 I

NSTRUCTION 

M

ANUAL V

. SB.3

1.  I

NTRODUCTION

• Channel 2

—the “auxiliary” DAC output, which con-

tinuously operates on Channel 1 for purposes of
limit evaluation, peak capture, tare offset, etc.  It is
produced for display, interrogation via the 

CHN

command, and generation of the “raw volts” analog
output (Channel 3).

• Channel 3

—the instrument’s “scaled voltage” out-

put, which is continuously proportional to Channel 2.
It may be set to a full scale of either ±5 or ±10 VDC,
and is available at the rear of the 3000PLUS as both
DC voltage and 4-20 mA output (see Section 2.D)

As explained in the following section, the front-panel

SCROLL

button may be used during normal run-time

operation to cycle through the three channel displays.

4

When Channel 2 or 3 is being displayed, the corre-
sponding front-panel indicator LED (

C2

or 

C3

) will

light—see Fig. 2.  No indicator lights when Channel 1 is
displayed.

The channel that is displayed whenever the 3000PLUS
is powered up is always the channel that was on dis-
play when the meter was last turned off. 

C

OMMUNICATIONS

I

NDICATION

Fig. 2 also shows the 3000PLUS instrument’s front-
panel 

CM

LED.  This indicator lights when the RS232

serial communications port (Section 2.B) is in receipt of
one or more transmitted ASCII characters (which may
or may not constitute a valid MNEMONIC COMMAND—
see Appendix A).  

L

IMIT

S

TATUS

I

NDICATION

Whenever continuous limit monitoring of the “auxiliary”
output (Channel 2) is enabled, the appropriate front-
panel limit status indicator (Fig. 2) will show the 

LIMIT

ZONE

in which Channel 2’s existing data reading lies

(regardless of whether Channel 2 is the one currently

D

ATA

D

ISPLAY

Updated four times a second, the 3000PLUS instru-
ment’s six-digit data LED data display is automatically
scaled to count by a resolution of 1, 2, or 5 to a maxi-
mum reading of ±199990.

1

Regardless of the scaling

currently in effect, a flashing display during normal run-
time operation indicates that an overrange condition
has occurred.  In this case, the displayed measurement
reading may be invalid.

During normal instrument setup, the operator will be
called upon to indicate the 

DESIRED FULL-SCALE

READING IN ENGINEERING UNITS

(the “FULL SCALE

UNITS” or “FSU” value), to correspond to a full-scale
output of 5.000 volts.  This can be done via the front-
panel setup procedure (as explained in Section 3.B) or
via the Configurator software, on a setup or run-time
basis  (as explained in Section 4).  The 

DECIMAL-

POINT RESOLUTION

of the data display for Channels

1 and 2 will always match that of the last-entered FSU
value.

2

You can also specify a 

DISPLAY OFFSET

value

(“DSO”), if desired—again, either through the front panel
or the Configurator software.  The DSO is a value of
numeric offset (positive or negative) to be continuously
applied to the displayed reading of Channel 1 or 2 (see
below).  Note that the DSO is applied in addition to any
zero offset resulting from instrument calibration (Sec-
tions 3 and 4) and to any specified tare offset (Section
5.D).

3

In addition to the normal-mode analog filtering fur-
nished by the installed 5D78 Conditioner, the 3000PLUS
can apply selectable 

DIGITAL FILTERING

(“DFL”) to the

data display to allow smooth, stable readout of dynamic
variables.  The DFL number ranges from 0 through 9 to
indicate increasing amounts of digital smoothing.  Like
the FSU and DSO values, it may be entered through
either the front panel or the Configurator software.

Use of the data display for manual setup of the
3000PLUS instrument is described in detail in Section 3
(“Front-Panel Configuration and Calibration”).

C

HANNEL

I

NDICATION

At any time, you can select any one of the 3000PLUS
instrument’s three separate output channels for display:

• Channel 1

—the meter’s basic ±5-VDC scaled out-

put, representing measured engineering units (after
calibration).  It is produced solely for display and
interrogation via the 

CHANNEL (CHN)

command

(described in Appendix A).

1.

D

D

ATA AND

S

TATUS

D

ISPLAYS

1

Under a reading of about 32000, the display’s least significant
digit will change by 1; from 32000 to 64000, it will change by 2;
and over 64000, by 5.  

2

That is, the precision (decimal-point location) both of the basic
±5-volt scaled output (Channel No. 1) and of the “auxiliary” DAC
output (Channel No. 2) will always reflect that of the currently
stored FSU number—as will that of other setup values that
directly relate to the 3000PLUS’s scaled engineering-units read-
ing, including display offset, high/low limit and hysteresis val-
ues, “peak defeat” threshold, tared output value, and all
calibration numbers (both “absolute” and “two-point”) that are
expressed in units.

3

Since the display offset is automatically set to zero during
instrument calibration (either “absolute” or “two-point”), you
should set a nonzero DSO value only after calibration has been
performed 
(Sections 3 and 4).

4

The Configurator software’s “Live Output Window” (described
in Section 4.D), lets you view any selected channel and modify
applicable output characteristics on a run-time basis.

Summary of Contents for 3000PLUS

Page 1: ...d reproduced or used in any form or by any electronic mechanical or other means including photocopying and recording or in any information storage and retrieval system without permission in writing fr...

Page 2: ...has been properly installed If your 3000PLUS is equipped with a 5D Series conditioner model other than the 5D78 the corresponding instruction manu al should be consulted ALWAYS TURN OFF THE 3000PLUS B...

Page 3: ...ront Panel Setup Procedure 3KP78 3 4 4 SOFTWARE CONFIGURATION AND CALIBRATION 4 A Using the 3000Plus Configurator Software 3KP78 4 1 4 B Summary of Configurator Menus 3KP78 4 3 4 C Overview of OFF LIN...

Page 4: ...ain Gage Cabling under 20 ft in length 3KP78 2 3 Fig 8 b 8 Wire Strain Gage Cabling 20 ft or longer 3KP78 2 3 Fig 9 Analog Output Connections 3KP78 2 4 Fig 10 3000PLUS Logic I O Connections Fig 10 a L...

Page 5: ...plained in Sections 3 and 4 of this manual you can quickly set up the 3000PLUS either via the simple front panel button menu or via the configuration soft ware supplied with the unit Operator entered...

Page 6: ...nd power supply communi cations terminals 1 INTRODUCTION High ESD immunity and extensive EMI protec tion further assure data integrity in harsh industrial environments Because of normal cable loading...

Page 7: ...inal 0 5 V where 5 V Logic 1 AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA A A A A 1 70 4 41 7 17 0 22 Fig 1 3000PLUS Dimensions true 25 V without damage noise im...

Page 8: ...d excitation frequency for an excitation of 10 00 kHz it is 2000 Hz for 5 00 kHz it is 1400 Hz and for 3 27 kHz it is 1100 Hz See Table 2 in Appendix B for the practical ranges that apply to the 5D78...

Page 9: ...ction 1 D Setup Indicators see Section 3 B Channel and Limit Value Indicators see Section 1 D Communi cations Indicator see Section 2 B Fig 2 3000PLUS Front Panel Elements Limit Logic Relay Outputs se...

Page 10: ...ser s Panel with Cutout and Mounting Holes see Fig 5 A B 4 060 3 700 0 180 0 200 0 156 dia Edge of 3000PLUS instrument 1 45 0 775 Fig 4 3000PLUS Panel Mounting Fig 5 Panel Cutout Dimensions 3 Remove t...

Page 11: ...setup or run time basis as explained in Section 4 The DECIMAL POINT RESOLUTION of the data display for Channels 1 and 2 will always match that of the last entered FSU value 2 You can also specify a DI...

Page 12: ...tification 2 RG RANGE input range and scaling information including for the 5D78 module excitation frequen cy full scale mV V range desired full scale reading in units and desired decimal point resolu...

Page 13: ...perator is called on to select one of a set of discrete values for a given setup parameter this button is used to cycle through the display of those values for details see Entering a Setup Parameter i...

Page 14: ...clears any currently captured peak value by momentarily disabling and re enabling peak capture see Section 5 B SETUP FUNCTIONS Used during setup to 1 Acknowledge each setup parameter when it appears...

Page 15: ...ut does require full installation of Microsoft Access 2000 Runtime which is supplied with the Configurator and which takes approximately 32 MB of hard drive space If Microsoft Office 2000 or higher is...

Page 16: ...Microsoft Office Office so that it reads C Program Files Microsoft Office Office11 f Click OK g Now you can run the Configurator by clicking on 3000PLUS Configurator in the Programs list THE CONFIGURA...

Page 17: ...want to completely remove the 3KP Configurator and all its components answer Yes to uninstall or No to abort You may be asked whether you re sure you want to remove MSCOMM32 OCX and COMDLG32 OCX which...

Page 18: ...r of the 3000PLUS instrument also shown in Fig 3 Each connected wire or jumper is to be firmly secured to the corresponding SCREW TERMINAL of the terminal block supplied with the meter that plugs into...

Page 19: ...nications between the 3000PLUS and an external PC While 3000PLUS PC serial com munications will usually take place through the 3KP Configurator software described in Section 4 a termi nal emulation pr...

Page 20: ...er 20 ft in length CAL SENSE SIGNAL CAL ENABLE EXCITATION SENSE SENSE EXCITATION SHIELD 10 9 8 7 6 5 4 3 2 1 SIGNAL EXTRA WIRE paired with CAL SENSE unconnected at Transducer Connector NOT ISOLATED co...

Reviews: