background image

3KP78 - 1.8

3000PLUS 

WITH 

5D78 I

NSTRUCTION 

M

ANUAL V

. SB.3

being displayed).  The three limit zones are defined by
the 3000PLUS instrument’s current 

HIGH LIMIT

(“HIL”)

and 

LOW LIMIT 

(“LOL”) settings:

• HI

(or “GREATER THAN”)—Channel 2’s reading is

greater than the current high limit value

• OK

(or “BETWEEN”)—Channel 2’s reading is greater

than or equal to the current low limit value and less
than or equal to the current high limit value

• LO

(or “LESS THAN”)—Channel 2’s reading is less

than the current low limit value

When limit monitoring is disabled, none of the limit indi-
cators will light.

For a complete discussion of limit monitoring, including
definition of limit setpoints, limits 

LATCH MODE

(“LAT”),

relay contact 

POLARITY

(“POL”), and both high-limit

and low-limit 

HYSTERESIS 

deadbands (“LHY” and

“HHY”), see Section 5.E.

R

UN

-T

IME

L

IMITS

D

ISPLAY

When the 3000PLUS instrument’s 

LIMITS SECURITY

(LMS)

is OFF, the local operator is able to use the front-

panel buttons (as explained in Sections 1.E and 5.E) to
quickly view and adjust the operating limit values during
normal run-time operation, without having to enter
Setup Mode
.  This is the case even when limit monitor-
ing is currently disabled (Section 5.E).

When limits security is ON, the operator can view and
modify the limits values for monitoring the “auxiliary”
DAC output only by following the standard front-panel
setup procedure given in Section 3, which may require
entry of a security code.

S

ETUP

S

TAGE

I

NDICATION

Each of the seven front-panel setup indicators shown in
Fig. 2 will light when the 3000PLUS enters the corre-
sponding stage of the 

FRONT-PANEL SETUP PROCE-

DURE

, as listed below and described in detail in

Section 3.  Except for 

ST

(“SETUP”), each indicator will

remain on only as long as the meter is in that setup
stage; 

ST 

will remain on until the operator exits setup

mode.

1.

ST

(“SETUP”)—security and module identification

2.

RG

(“RANGE”)—input range and scaling information,

including (for the 5D78 module) excitation frequen-
cy, full-scale mV/V range, desired full-scale reading
in units, and desired decimal-point resolution

3.

FL 

(“FILTER”)—analog and digital filter settings

4.

CL

(“CALIBRATION”)—including desired calibration

method and all cal-point values applicable to that
method

5.

LM

(“LIMIT”)—parameters relating to the limit moni-

toring of the “auxiliary” output (Channel 2), including
limit enable, latch mode, relay polarity, setpoint val-
ues, and hysteresis windows*

6.

PK

(“PEAK”)—parameters relating to the further pro-

cessing of the “auxiliary” output (Channel 2), includ-
ing peak mode, “defeat” threshold, “backout”
threshold, decay rate, and tare offset

7.

AN

(“ANALOG OUTPUT”)—full-scale voltage output,

followed by entry a new security code, if desired

M

ODULE

S

TATUS

I

NDICATION

Shown in Fig. 3, the status indicator light of the installed
5D78 Conditioner module serves to monitor the mod-
ule’s power, input, communications, and general health
condition.  The condition(s) represented by the light’s
three possible colors and color combinations are given
below.  

NOTE: 

As long as the module is properly communicat-

ing with the 3000PLUS meter (and the meter is pow-
ered up), the indicator light will be flashing

if the flashing light is constantly 

GREEN

, the mod-

ule’s input signal is OK

if the flashing light is constantly 

YELLOW

, the mod-

ule’s input signal is over 20% out of range

if the flashing light is constantly 

RED

, a serious input

condition has been detected (e.g., excessive cur-
rent, overvoltage); it could indicate a transducer
short or faulty cabling

if the flashing light is alternating 

YELLOW AND

GREEN

, the module has received a mnemonic

command from the 3000PLUS meter (the yellow
light will continue for about a second after receipt of
the command-terminating carriage return)

if the flashing light is alternating 

RED AND GREEN

,

a significant internal software error detected; contact
the Daytronic Service Department

1.  I

NTRODUCTION

* Also on when viewing limits in run-time (see above).

Summary of Contents for 3000PLUS

Page 1: ...d reproduced or used in any form or by any electronic mechanical or other means including photocopying and recording or in any information storage and retrieval system without permission in writing fr...

Page 2: ...has been properly installed If your 3000PLUS is equipped with a 5D Series conditioner model other than the 5D78 the corresponding instruction manu al should be consulted ALWAYS TURN OFF THE 3000PLUS B...

Page 3: ...ront Panel Setup Procedure 3KP78 3 4 4 SOFTWARE CONFIGURATION AND CALIBRATION 4 A Using the 3000Plus Configurator Software 3KP78 4 1 4 B Summary of Configurator Menus 3KP78 4 3 4 C Overview of OFF LIN...

Page 4: ...ain Gage Cabling under 20 ft in length 3KP78 2 3 Fig 8 b 8 Wire Strain Gage Cabling 20 ft or longer 3KP78 2 3 Fig 9 Analog Output Connections 3KP78 2 4 Fig 10 3000PLUS Logic I O Connections Fig 10 a L...

Page 5: ...plained in Sections 3 and 4 of this manual you can quickly set up the 3000PLUS either via the simple front panel button menu or via the configuration soft ware supplied with the unit Operator entered...

Page 6: ...nd power supply communi cations terminals 1 INTRODUCTION High ESD immunity and extensive EMI protec tion further assure data integrity in harsh industrial environments Because of normal cable loading...

Page 7: ...inal 0 5 V where 5 V Logic 1 AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA AAAAAAAAAAAAA A A A A 1 70 4 41 7 17 0 22 Fig 1 3000PLUS Dimensions true 25 V without damage noise im...

Page 8: ...d excitation frequency for an excitation of 10 00 kHz it is 2000 Hz for 5 00 kHz it is 1400 Hz and for 3 27 kHz it is 1100 Hz See Table 2 in Appendix B for the practical ranges that apply to the 5D78...

Page 9: ...ction 1 D Setup Indicators see Section 3 B Channel and Limit Value Indicators see Section 1 D Communi cations Indicator see Section 2 B Fig 2 3000PLUS Front Panel Elements Limit Logic Relay Outputs se...

Page 10: ...ser s Panel with Cutout and Mounting Holes see Fig 5 A B 4 060 3 700 0 180 0 200 0 156 dia Edge of 3000PLUS instrument 1 45 0 775 Fig 4 3000PLUS Panel Mounting Fig 5 Panel Cutout Dimensions 3 Remove t...

Page 11: ...setup or run time basis as explained in Section 4 The DECIMAL POINT RESOLUTION of the data display for Channels 1 and 2 will always match that of the last entered FSU value 2 You can also specify a DI...

Page 12: ...tification 2 RG RANGE input range and scaling information including for the 5D78 module excitation frequen cy full scale mV V range desired full scale reading in units and desired decimal point resolu...

Page 13: ...perator is called on to select one of a set of discrete values for a given setup parameter this button is used to cycle through the display of those values for details see Entering a Setup Parameter i...

Page 14: ...clears any currently captured peak value by momentarily disabling and re enabling peak capture see Section 5 B SETUP FUNCTIONS Used during setup to 1 Acknowledge each setup parameter when it appears...

Page 15: ...ut does require full installation of Microsoft Access 2000 Runtime which is supplied with the Configurator and which takes approximately 32 MB of hard drive space If Microsoft Office 2000 or higher is...

Page 16: ...Microsoft Office Office so that it reads C Program Files Microsoft Office Office11 f Click OK g Now you can run the Configurator by clicking on 3000PLUS Configurator in the Programs list THE CONFIGURA...

Page 17: ...want to completely remove the 3KP Configurator and all its components answer Yes to uninstall or No to abort You may be asked whether you re sure you want to remove MSCOMM32 OCX and COMDLG32 OCX which...

Page 18: ...r of the 3000PLUS instrument also shown in Fig 3 Each connected wire or jumper is to be firmly secured to the corresponding SCREW TERMINAL of the terminal block supplied with the meter that plugs into...

Page 19: ...nications between the 3000PLUS and an external PC While 3000PLUS PC serial com munications will usually take place through the 3KP Configurator software described in Section 4 a termi nal emulation pr...

Page 20: ...er 20 ft in length CAL SENSE SIGNAL CAL ENABLE EXCITATION SENSE SENSE EXCITATION SHIELD 10 9 8 7 6 5 4 3 2 1 SIGNAL EXTRA WIRE paired with CAL SENSE unconnected at Transducer Connector NOT ISOLATED co...

Reviews: