Chapter 3 - How to...
3.4.8.3 Cleaning
CAUTION
Wear powder-free gloves at all times when cleaning the device including the scan head and
main mirror. If the gloves become contaminated they must be replaced immediately.
In order to prevent any contamination always clean and disinfect parts of the equipment
which will be in contact with the patient, see
page.
Adequate cleaning and disinfection is necessary to prevent disease transmission.
Thoroughly clean and disinfect the equipment surfaces that contact the patient and all
equipment surfaces likely to become soiled during use.
Improper cleaning methods or use of inappropriate cleaning and disinfecting agents can
result in the following:
l
Damaged equipment
l
Poor imaging performance
l
Increased risk of electric shock
Always follow the cleaning and disinfecting agents’ manufacturer’s instructions and
precautions; for storage, method of application, protective gloves, shelf life and disposal to
help assure effective and safe use of the product.
Do
NOT
use lint cloths, tissues, or other materials that may create dust, on or near the scan
head. Always allow cleaned areas to air-dry.
Do
NOT
use harsh detergent, abrasive cleaners or solvents that are
NOT
recommended by
the manufacturer.
Do
NOT
expose equipment parts to steam or sterilizers as this will damage the device.
Do
NOT
allow liquids to get into the equipment.
If you believe cleaning sprays or liquids have got into the equipment, disconnect the power
supply and have the equipment checked by Optos qualified service personnel.
Do
NOT
let the cleaning wipes come into contact with the inside of the instrument.
Do
NOT
use the equipment until you have confirmed it is safe to use.
3.4.8.3.1 General cleaning
CAUTION
Be careful
NOT
to damage the scan head covers when cleaning.
The equipment should be kept clean and dust free.
l
Logout, shut down and power off the device before cleaning the outer casing of the scan head.
l
Use a soft, slightly damp cloth to clean the plastic surfaces.
l
Ensure debris does
NOT
fall inside the device.
l
Use a glass-cleaning agent to clean the PC monitor and touch screen.
Page 40 of 60
English
Part
Number:
G102748/11_ENG
Copyright
2021
,
Optos
plc.
All
rights
reserved.
Summary of Contents for P200T
Page 1: ...optos com User Guide P200T Optos Panoramic Ophthalmoscope G102748 11_ENG...
Page 4: ...Page 4 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 10: ...Page 10 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 26: ...Page 26 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 58: ...Page 58 of 60 English Part Number G102748 11_ENG Copyright 2021 Optos plc All rights reserved...
Page 60: ...G102748 11_ENG English...