background image

13

Step 9:

  Using a soft, lint-free cloth or towel wipe the sleeve down using a 

commercial scale cleaner (i.e. CLR

®

 or similar). This removes scaling or iron 

deposits that may be on the outside of the quartz sleeve. Be careful not to 

get any moisture or liquids inside of the sleeve.

Step 10:

 Dry the sleeve with separate cloth.

Step 11:

 Replace the o-ring and slide the sleeve back into the chamber following steps 

7 and 8 from the installation section of the manual.

Cleaning the UV Sensor

Depending on the water quality, the UV sensor may require periodic cleaning. 

At a minimum, the UV sensor should be cleaned on an annual basis. The following 
steps outline a basic cleaning procedure.

Step 1:

  If a by-pass assembly is installed, shut the inlet valve off to prevent water 

flow through the system. Otherwise, turn off main water inlet valve (and/or 

turn off the water pump).

Step 2:

  Disconnect power cord of UV system from electrical outlet.

Step 3:

  Release water pressure by opening a downstream faucet and then close 

the outlet shut-off valve (if any). If there is no outlet shut-off valve, expect 

water to drain from the system as the head pressure in the system will 

cause the water to flow back down.

Step 4:

  Place something under the chamber to catch any water that may come out 

of the chamber during the removal of the UV sensor.

Step 5:

  Unscrew (counterclockwise) sensor nut from the chamber and pull the 

sensor slowly out of the sensor port.

Step 6:

  Holding the sensor in your hand wipe the flat portion (sensor face) of the 

sensor with isopropyl alcohol using a clean lint-free cloth.

Step 7:

  Replace sensor following step 9 from the installation section of the manual.

Operation

Microlene systems come with a feature laden controller that incorporates both 

the lamp driver (ballast) and control features in one water-tight case. Two main 

controllers are available for the Microlene systems (depending on your model). 

Both models feature a power factor corrected, constant current lamp driver with 

a universal power input. 
Please Note: While the LED or display screen is red and the buzzer is sounding 

the water from the system should NOT be consumed. If any water does pass 

through the system during this period, please follow the disinfection procedure as 

outlined in this manual before the water is consumed. For Microlene systems, even 

though they have a visual and audible warning built into the controller, a green 

LED or status screen does not necessarily indicate that the water coming from 

this system is in fact potable (safe to drink). These systems do not measure the 

level of disinfection; they simply measure the “on-off” status of the lamp. Please 

have your water checked for microbiological contaminants on a regular basis.

IMPORTANT: Installation of a backflow valve should be in compliance with 

all local water authority regulations and in accordance with AS/NZS 3500.1.2

and complying with AS/NZS 2845.1.

Summary of Contents for microlene MCS

Page 1: ...he operator of this equipment Aquashield Centurion 3 Stage UV Disinfection System Models MCS MCXS without cartridges KMCS KMCXS with cartridges included Installation and Operating Instructions Flow Rate 30mJ cm2 95 UVT 57 L min AS NZS 3497 ...

Page 2: ... reassurance that your family will have access to safe drinking water throughout your entire home protecting from microbiological contamination This is a chemical free process which is simple in its concept and effective in its abilities to inactivate microorganisms present in the water supply Simple maintenance Continuous disinfection Safe water Microlene makes it that easy These instructions are...

Page 3: ...Operation 13 MCS Controllers 14 MCXS Controllers 14 MCXS Power up Sequence 14 MCXS Operational Screens 15 MCXS with optional sensor Operational Screens 16 Lamp Countdown Sequence 16 Lamp Replacement MCS systems 17 Lamp Replacement MCXS systems 17 QR Codes 17 System Troubleshooting 18 Temperature Management Devices 19 Expansion Modules 20 Microlene Aquashield Centurion UV System Specifications 21 L...

Page 4: ...l capabilities or lack of experience and knowledge unless they have been given supervision or instruction 6 Children should be supervised so that they do not play with the appliance 7 To avoid possible electrical shock use only with a properly earthed electrical outlet Appliances that are obviously damaged must not be operated 8 Never perform any maintenance to the system unless you are comfortabl...

Page 5: ...erials Inlet outlet connections Teflon thread tape Water Quality Parameters UV disinfection is extremely effective against microorganisms but only if the UV light can pass through the water it needs to treat This means that the quality of your water is very important in order to ensure complete disinfection Treated water should be tested for at least the parameters listed below If the water exceed...

Page 6: ... MC HOSL 7 UV Chamber 8 Clamps 9 Flexible Hose 10 Mounting Plate 11 Filter Cartridges Only KMCS and KMCXS have filters included Part 20PP20J 20 20 micron sediment Part 1PS20J 20 1 micron sediment 12 Filter Housings Part DM160018 double filter housing two 20 sumps and caps 13 Mounting Screws 14 Filter Wrench Part WFH45 15 UV Sensor optional module Part MCXS UVS 16 IEC Power Cable Part MC CONT PC Au...

Page 7: ...or above the chamber IMPORTANT All Microlene Aquashield Centurion UV Disinfection Systems are intended to be installed under a weather proof cover and should not be exposed to the elements MCS C Weather proof cover purchased separately Protects the chamber and controller from rain and direct elements Enables the 3 Stage UV system to be set up left to right or right to left for ease of installation...

Page 8: ...cating water flow which now should be indicating a flow direction of right to left Figure 2 System Orientation water inlet Step 4 Remove the stainless steel UV chamber from the two plastic clamps located on the top of the rack Carefully remove the top straps securing the chamber with the aid of a standard slot screwdriver Rotate the chamber 180 degrees with the inlet now facing to the left and the...

Page 9: ...gently remove the quartz sleeve from its packaging being careful not to touch the sleeve or lamp with your hands The use of cotton gloves is recommended for this procedure as oils from the hands can leave residue on the sleeve and lamp which can ultimately block the UV light from getting to the water Carefully slide the sleeve into the chamber until you can feel it hit the opposite end of the cham...

Page 10: ... controller Align the flat portion so it faces the gland nut end and matches up with the half metal lip on the sensor port see Figure 6 Insert the sensor so it is fully seated and hand tighten the sensor nut Insert the sensor connector into the IEP port located on the right side of the controller Figure 7 For the sensor to be recognized by the controller the controller power must be plugged in las...

Page 11: ... lamp Note the keying for proper alignment see Figure 8 Insert the lamp connector into the gland nut and turn the connector approximately turn to lock the connector to the gland nut as in Figure 10 Figure 9 Lamp Key Installation Figure 10 Lamp Connector Step 12 Tighten the captive earth screw to the earth lug on the UV chamber to ensure proper earthing Figure 11 Earth Screw Connection Step 13 Your...

Page 12: ...ng Step 4 Reinstall the filter cartridge into the sump and flush the chlorine solution by opening all faucets until chlorine can no longer be detected Your home has now been completely disinfected with your Microlene Aquashield Centurion UV system ready to inactivate any microorganisms that enter the home Cleaning the Quartz Sleeve Depending on the water quality the quartz sleeve may require perio...

Page 13: ...ise sensor nut from the chamber and pull the sensor slowly out of the sensor port Step 6 Holding the sensor in your hand wipe the flat portion sensor face of the sensor with isopropyl alcohol using a clean lint free cloth Step 7 Replace sensor following step 9 from the installation section of the manual Operation Microlene systems come with a feature laden controller that incorporates both the lam...

Page 14: ...ription of the system s performance in addition to providing any applicable fault messages and system diagnostics The controller can be used as both a non monitored and UV monitored system If a true UV monitor is required simply add the optional UV sensor module to the infinite expandability port located on the right side of the controller Simply plug in an optional UV sensor module into the expan...

Page 15: ... ready to disinfect water flow all detected modules lamp reaching max output successful start up MCXS Operational Screens On systems without the UV monitor the default screen shows the Microlene Home Screen At any point during operation the user is able to scroll through the Microlene Home Screen Lamp life remaining QR Code Contact Info and Maintenance Parts screens by pressing the button located ...

Page 16: ...tant audible alarm is given This alternates with a screen indicating water may be unsafe for consumption With the solenoid module the controller de activates the solenoid valve shutting off all water flow audible chirp every 15 seconds audible chirp every 15 seconds constant audible alarm cycles with red low uv screen Lamp Countdown Sequence The system counts down the number of days until a lamp c...

Page 17: ...with step 11 page 11 for instructions on installing the new lamp To reset the timer in the controller firmly hold down the button on the controller for 10 seconds The controller will read rSt3 rSt2 rSt1 and then beep The button can now be released the lamp countdown timer has been reset The lamp key that comes with the replacement lamp will not be required on the MCS system and can be discarded La...

Page 18: ... the chamber or on the Maintenance parts screen Low UV Intensity Remove and clean the quartz sleeve and sensor Check water quality meets requirements on page 5 and add filtration as required Replace lamp Wrong lamp or sensor installed Replace component with proper model as indicated The UV sensor is no longer communicating with the system Ensure all modules are connected properly to the system and...

Page 19: ... heading Boil Water Advisory If any failure occurs on a Microlene UV system the water must not be used for human consumption until the system is returned to a safe operational mode If the water is used for human consumption during this period the water must be boiled minimum 20 minutes at a full boil prior to consumption Temperature Management Devices Your Microlene system is designed to run conti...

Page 20: ... of contacts In normal operation the OK and COM contacts will be connected and in a fault condition Low UV Lamp fail Power Fail the Fault and COM contacts will be connected Maximum Contact Rating is 30V 1A use 16 22 AWG Order PN DMMOD RAM SOLENOID CONNECTION MODULE Connects a NORMALLY CLOSED line voltage solenoid valve to the controller On a non monitored system the solenoid will only close on a l...

Page 21: ...lamp Power Watts 57 48 230V Replacement Lamp MC HOLP Replacement Sleeve MC HOSL Chamber Dimensions 8 9 x 50 8 cm 3 5 x 20 Chamber Material Polished 316L Stainless Steel A249 Pressure Rated Tubing Controller Dimensions 21 8 x 10 7 x 8 9 cm 8 6 x 4 2 x 3 5 Operating Pressure 827kPa 120PSI Operating Water Temperature 2 40 C 36 104 F UV Monitor not available OPTIONAL optional UV module sold separately...

Page 22: ...22 NOTES ...

Page 23: ...23 ...

Page 24: ...of date of original purchase If requested return the product and or provide further information with respect to the claim Returning the product to the place of purchase is at your cost and is your responsibility The warranty claim will be assessed by Davey on the basis of their product knowledge and reasonable judgement and will be accepted if a relevant defect is found the warranty claim is made ...

Reviews: