1.2 Data Collection Specifications
Temperature:
• Measurement Range:
25 °C - 50 °C
• Accuracy:
Ta: ± 0.05 °C
• Resolution:
0.01 °C
• Convection:
5 ‘puck’ sensors (T6 - T10) - radiant warmer
• Air Conduction:
5 sensors (alternate ‘post’ Ta - Te) - incubator
1 mattress temperature
• Optional Skin Sensor
‘oven’ with thermal pad
contact:
28 °C to 40 °C in 2 °C increments ± 0.01 °C resolution,
± 0.05 °C Accuracy
Airflow:
• Measurement Range:
0.1 to 1 m/s
• Accuracy:
± 0.1 m/s
at 50% RH ± 15%
Relative Humidity:
• Measurement Range:
0 - 100%
Non Condensing
• Accuracy:
± 3 %
• Resolution:
0.1% RH
Sound Level:
• Measurement Range:
30 dbA to 100 dbA, Type III
• Accuracy:
± 3 dbA
• Resolution:
0.1 dbA
Data Retention:
• Capacity:
up to 48 hours per test
• Sample Rate:
1/min or 1/sec, depending on parameter
Specifications/Chapter 1
#