vPad-TI OPERATING MANUAL
If more than one temperature has been selected on the Autosequence, the test will begin with the
lowest temperature value defined in the test. When the test starts, the screen will change to the
Temperature Stabilization screen indicated in Figure 46. While this screen is displayed, the system is
waiting for the Oven temperature to operate consistently within ± 0.1 °C. To abort the
Skin
Sensor Accuracy
test click on the
Cancel
button.
Once the Oven temperature has stabilized at the set temperature, the program will automatically
proceed to the Sensor Temperature Input screen, Figure 47. On this screen the temperature of the
Skin Sensor, as shown by the Control Panel of the incubator, can be entered. If another test (air
temp, warmup, overshoot, etc.) is in progress when the
Probe Test
requests the Sensor
temperature, the incubator can be switched between Air control and Skin control
momentarily
to
read the Skin Temperature, without affecting the
Air Temperature Control
test.
Figure 45
Figure 46
: Oven Temperature Stabilization
Running a Test/Chapter 5
#
Page 40