![Datex-Ohmeda S/5 M-EEG Technical Reference Manual Slot Download Page 7](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-eeg/s-5-m-eeg_technical-reference-manual-slot_513842007.webp)
S/5 EEG Module, M-EEG and S/5 Headbox, N-EEG
Document No. 800 1011-1
3
1.2.2
AEP
Amplification
10 000
Resolution
60 nV
Max amplitude
1000
µ
V
pp
Stimulation
Click (condensating)
duration 100
µ
s
Frequency
1.1...9.1 Hz (1 Hz steps) @ 10 ms measurement
1.1...8.1 Hz (1 Hz steps) @ 100 ms measurement
Intensity
10...90 dB nHL, 10 dB steps
Measurement
Sampling frequency
2400 Hz for MLAEP/ 4800 for BAEP
Frequency range
0.5...1000 Hz
Highpass filter
off/10/30/50/75/100/150 Hz
Single average:
Averaged responses
100...2000 stimuli
Moving average:
Gross average
100...2000 stimuli
Update interval
after every 100 stimuli (200, when gross average is 2000)
1.2.3
EMG
Amplification
50 000
Resolution
100 nV
Max amplitude
100
µ
V
pp
Frequency range
60...300 Hz
Amplitude
Root Mean Square (RMS)
1.2.4
Impedance measurement
Measurement frequency
75 Hz
Current
10
µ
A
Range
0...30 k
Ω
Resolution
100
Ω
Accuracy
±
1k or
±
10 % whichever is greater
Measurement time, all leads 3 s
Start of measurement
manual/automatic
Leads off detection
>3 M
Ω
, continuous