![Datex-Ohmeda S/5 M-EEG Technical Reference Manual Slot Download Page 25](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-eeg/s-5-m-eeg_technical-reference-manual-slot_513842025.webp)
S/5 EEG Module, M-EEG and S/5 Headbox, N-EEG
Document No. 800 1011-1
21
BSR
shows the burst-suppression classification of EEG waveforms. 0 means artifact, 1
suppression, 2 burst and 3 invalid (the EEG is not classified).
Artefact
shows the artefact status of all channels. 0 means that there are no artefacts on the signal
and 1 means that there are artefatcs.
Noise
shows 1 if the activity of FEMG is too high for clean EEG signal and that EEG is not probably
reliable.
EMG AD
shows the latest AD conversion result of FEMG signal without any filtering or scaling.
Leadset AD
shows the latest AD conversion result of lead set signal without any scaling.
Bipolar
shows the montage type that is active in module and headbox. 1 means bipolar and 0
referenced.
60 Hz
is 1 if the power frequency filter is set for 60 Hz power frequency.
Imp. Meas.
shows 1 if the impedance measurement is on in the headbox.
Imp. Check Failed
shows 1 if the impedance check has failed.
EP Meas.
shows 1 if the evoked potential measurement is active in headbox.
High EP Sampl.
Rate
shows 1 if the headbox uses higher sampling rate for 10 ms EP
measurement.
EP Epoch Points
shows the number of samples collected for each epoch.
EP Channels
shows the number of channels used in EP measurement.
Hb ROM Error
is 1 if the checksum at the EPROM is not accordance with the one the software has
calculated.
Setup
The items in setup menu are collection of items in normal EEG and EP menus. There are no special
service settings in this menu.
Imp. Cycle
Impedance measurement repetition time.
Select Montage
Selection of headbox’s electrode configuration.