![Datex-Ohmeda S/5 M-EEG Technical Reference Manual Slot Download Page 18](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-eeg/s-5-m-eeg_technical-reference-manual-slot_513842018.webp)
Datex-Ohmeda S/5 monitors
Document No. 800 1011-1
14
3.3
Disassembly and reassembly
3.3.1
M-EEG
Disassemble the M-EEG in the following way. See
Exploded view of module box and
Remove the two screws from the back of the module.
2.
Pull the module box slowly rearward and detach it from main body. Be careful with the loose
latch and spring locking pin.
3.
Detach the Neuro board by removing the two screws located near the front panel frame,
disconnect the cable and pull out the front panel frame.
To reassemble the module, reverse the order of the disassembly steps.
CAUTION
When reassembling the module, make sure that the cables are reconnected
properly.
3.3.2
N-EEG
Disassemble the N-EEG in the following way. See
1.
Remove the four screws from the bottom of the headbox.
2.
Lift off the cover and disconnect the two cables connected to the EEG headbox board.
3.
Disconnect the module-headbox cable and the headbox input unit connectors from the EEG
headbox board.
4.
Remove the four screws on the corners of the EEG headbox board and detach the EEG
headbox board.
To reassemble the N-EEG, reverse the order of the disassembly steps.
CAUTION
When reassembling the headbox, make sure that the cables are reconnected
properly.