Datex-Ohmeda S/5 monitors
Document No. 800 1011-1
16
4.2
Troubleshooting flowcharts
Possibly faulty EEG
module.
Take all modules out of
the frame.
Insert M-EEG and turn
power on
Does fault
still appear?
Fault not in EEG module
Enter the EEG Service
menu
OK?
Does another
module work in
the same slot?
No
No
No
Yes
Check are the Timeouts,
Bad checksums or Bad
c-s by mod running?
Check RAM, ROM and
EEPROM
OK?
Replace Neuro board
Yes
No
Connect headbox into
module connector
Yes
Yes
Check from headbox
view are the Timeouts,
HB Mod Bad
Checksums or Mod Hb
Bad Checksums running
OK?
Check front panel key
functions
Yes
Does another
module work with
headbox?
No
Yes
Fault not in EEG module
No
OK?
Is the keypad
connected?
No
Connect the keypad
No
Does the test
pass with
another
keypad?
Yes
Replace the keypad
Yes
Replace Neuro board
No
Connect earphones into
the headbox and start
AEP measurement
Yes
Can clicks
be heard?
No
Can clicks be heard
with another
headbox?
Yes
Fault not in EEG module
No
WHITE
Figure 5
EEG module troubleshooting flowchart