S/5 EEG Module, M-EEG and S/5 Headbox, N-EEG
Document No. 800 1011-1
17
Possibly faulty EEG
headbox.
Take all modules out of
the frame.
Insert M-EEG and EEG
headbox and turn power
on
Does
fault still
appear?
Fault not in EEG
headbox
Enter the EEG Service
menu
Does the
headbox work
with another
module?
No
Check from headbox
view are the Timeouts,
HB Mod Bad
Checksums or Mod Hb
Bad Checksums running
OK?
Check front panel key
functions
Yes
Fault in cable?
No
Replace the headbox -
module cable
OK?
Is the keypad
connected?
No
Connect the keypad
No
Does the test
pass with
another
keypad?
Yes
Replace the keypad
Yes
Replace Neuro board
No
Connect earphones into
the headbox and start
AEP measurement
Yes
Can clicks
be heard?
No
Yes
Yes
No
Try with another
headbox - module cable
or check the cable by
hand
Replace the Headbox
board
No
Yes
Check HB Rom Error
OK?
No
Yes
Try with another
headbox - module cable
or check the cable by
hand
Fault in cable?
Replace the headbox -
module cable
Yes
Is earphone jack intact
and connected to
headbox circuit board?
No
Connect or replace the
earphones jack
No
Yes
Yes
Check EEG waveforms,
FEMG reading,
impedance
measurement and lead
off detection with EEG
simulator
OK?
Replace the Headbox
board
Check EP waveforms
with EEG simulator
OK?
Yes
No
Try with another
headbox - module cable
or check the cable by
hand
Fault in cable?
Replace Neuro board
No
Yes
Check the headbox input
connector and input
board
OK?
Yes
Replace the input
connector
No
Text
Figure 6
Headbox troubleshooting flowchart