![Datex-Ohmeda S/5 M-EEG Technical Reference Manual Slot Download Page 24](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-eeg/s-5-m-eeg_technical-reference-manual-slot_513842024.webp)
Datex-Ohmeda S/5 monitors
Document No. 800 1011-1
20
HB Mod Timeouts
is a cumulative number that indicates how many times the headbox has not
responded to module’s inquiry.
HB Mod Bad Checksum
is a cumulative number that indicates how many times there has been an
error in the message from headbox to module.
Mod HB Bad Checksum
is 1 if there has been errors in the messages from module to headbox.
Mod Mon Timeouts
is a cumulative number that indicates how many times the module has not
responded to monitor’s inquiry.
Mod Mon Bad Checksum
is a cumulative number that indicates how many times there has been
an error in the message from module to monitor.
Mon Mod Bad Checksum
is a cumulative number that indicates how many times there has been
an error in the message from monitor to module.
RAM
indicates the state of the RAM memory.
ROM
indicates whether the checksum at the EPROM is accordance with the one the software has
calculated.
EEPROM
indicates if the values stored in permanent memory are valid.
The states in memory checks are
OK, Fail
or
?
(module not in place or a communication error).
Headbox
Service data is changed to headbox view.
EP Start/Stop
is highlighted when the EP Start/Stop button is pressed in headbox or module.
Imp. Check
is highlighted when the Imp. Check button is pressed in headbox or module.
Active
shows 1 if the channel is active.
Lead off+
shows the lead off status of all plus electrodes. 0 means that the lead is on and 1 that
the lead is off.
Lead off-
shows the lead off status of all minus electrodes. 0 means that the lead is on and 1 that
the lead is off.
Imped+
shows the impedance of plus electrodes of last impedance check in all channels.
Imped-
shows the impedance of minus electrodes of last impedance check in all channels.