![Datex-Ohmeda S/5 M-COPSv Technical Reference Manual Slot Download Page 6](http://html1.mh-extra.com/html/datex-ohmeda/s-5-m-copsv/s-5-m-copsv_technical-reference-manual-slot_3795269006.webp)
Datex-Ohmeda S/5 monitors
Document No. 800 1012-1
2
1
SPECIFICATIONS
1.1
General specifications
Module size (W
×
D
×
H)
37
×
180
×
112 mm / 1.5
×
7.1
×
4.4 in
Module weight
0.35 kg / 0.8 lbs
Power consumption, M-COP
Approximately 3.5 W
Power consumption, M-COPSv
Approximately 5 W
1.2
Typical performance
1.2.1
C.O.
Measurement range
0.1...20 l/min
Display resolution
0.01 l/min
Repeatability
±
2 % or
±
0.02 l/min whichever is greater
(measured from electrically generated flow curves)
Max. change in blood temp
2.99 °C
Injectate temp range (with Edward's compatible probes)
0...25.5 °C
±
0.3 °C
25.5...27.0 °C
±
0.5 °C
Blood temp range (with Edward's compatible catheters)
17.5...30.9 °C
±
0.5°C
31.0...43.0 °C
±
0.3°C
Protection against electric shock
type CF defibrillation proof
REF
Repeatability
±
2 %
(Measuring range 10-60 %)
1.2.2
SvO
2
Accuracy
±
2 %
(Measuring range 30-95 %)
Equal to standard deviation when using in-vivo calibration.
1.2.3
InvBP
Measurement range
-40...+320 mmHg
Zero adjustment range
±
150 mmHg
Calibration range
±
20 %