Datex-Ohmeda S/5 monitors
Document No. 800 1012-1
26
5.1
COP Menu
COP
Record Data
Record Data
prints out the service data and module information (id. serial number and software
id.) on the Recorder Module, M-REC.
Service Data
P4
Gain
is a coefficient to compensate for gain error. Typically the value is between 17000 and
25000. Calibrate if the zero and/or gain value is outside the ranges.
Zero
indicates the offset compensation value of each parameter in the A/D converter. Usually the
value is within
±
1000.
Cable
shows ON when the corresponding cable is connected to the front panel and
Probe
shows
ON when the corresponding probe is connected to the cable.
Value shows
the measured numeric values simultaneously. Pressure value is a real time value and
shown in mmHg.
Probe
items
Catheter
(ON/OFF) and
Inj.
(FT, BATH, or OFF) indicate connections and
Value
indicates the measured temperatures in 0.01
°
C increments.
SvO2
Meas. state:
Measurement status shows: No optical module (No OM) connected, initializing the
optical module (Init OM), normal measurement state (Normal) and failed module (OM fail).
Value
is a measured SvO
2
value.
Gain
is the gain of the remote red and infrared signals (0, 1, 2 or 3)
Red int
: Reflected red intensity
Ired int
: Reflected infrared intensity
Loc red
: Local red intensity