5.1.1
COP calibration menu
Calibrate P4
This function is for calibrating the invasive blood pressure channel P4.
The calibration requires a pressure transducer (with an appropriate cable) and a pressure
manometer.
Calibration:
1.
Connect the pressure transducer with the pressure manometer to the P4 connector. Select
Calibrate P4. Leave the transducer at room air pressure.
2.
Press the ComWheel to start zeroing.
3.
Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
4.
Set the pressure on the display to match the pressure reading on the manometer and press
the ComWheel.