Cardiac Output Modules, M-COP and M-COPSv
Document No. 8001012-1
15
3.2
Service check
These instructions include complete procedures for a service check. The service check is
recommended to be performed after any service repair. However, the service check procedures can
also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (
Appendix A
) which should be filled in when performing the
procedures.
The mark
?
in the instructions means that the check form should be signed after performing
the procedure.
The procedures are designed for monitors with S/5 monitor software of revision 01. However, most
of the procedures also apply to monitors, which contain some other monitor software
type/revision.
3.2.1
Recommended tools
Tool
Order No.
Notes
Patient simulator
SvO
2
simulator
890121
Pressure manometer
InvBP transducer
Catheter connecting cable
Screwdriver
All modules
•
Detach the module box by removing the two screws from the back of the module. Be careful
with the loose latch and spring locking pin.
1.
Check internal parts:
•
screws are tightened properly
•
cables are connected properly
•
all socket mounted IC’s are inserted properly
•
EMC covers are attached properly
•
there are no loose objects inside the module
?
2.
Check external parts: