![Datex-Ohmeda S/5 M-COPSv Technical Reference Manual Slot Download Page 24](http://html1.mh-extra.com/html/datex-ohmeda/s-5-m-copsv/s-5-m-copsv_technical-reference-manual-slot_3795269024.webp)
Datex-Ohmeda S/5 monitors
Document No. 800 1012-1
20
?
•
Fill in all necessary documents.
3.3
Disassembly and reassembly
Disassemble the Cardiac Output Modules ,M-COP and M-COPSv, in the following way. (see the
exploded view of the module.
1.
Remove the two screws from the back of the module.
2.
Pull the module box slowly rearwards and remove it from main body. Be careful with the
loose latch and spring locking pin.
To reassemble the module, reverse the order of the disassembly steps.
CAUTION
When reassembling the module, make sure that the cables are reconnected
properly.
3.4
Adjustments and calibrations
3.4.1
Cardiac output calibration
The cardiac output calibration can be performed only at the factory.
3.4.2
Invasive pressure calibration
Calibrate invasive pressure when the pressure transducer (probe) is replaced with a different type
of transducer.
1.
Enter the COP service menu (
Monitor Setup
- Install/Service - Service - Parameters
)
2.
Enter the
Calibrations
menu.
3.
Connect a pressure transducer with a pressure manometer to the P4 connector. Select