Datex-Ohmeda S/5 monitors
Document No. 800 1009-1
30
All modules
Turn the monitor off, disconnect the Gas interface cable and reassemble the module. Remember to
attach the plastic cover against the CPU board before installing the module box.
NOTE: When reassembling the module make sure that the tubes are not pinched between the
module box and internal parts.
23.
Perform electrical safety check and leakage current test.
?
Install the Compact Airway Module into the Central Unit, turn the monitor on and wait until the
message ‘Calibrating gas sensor’ disappears from the screen.
24.
Block the tip of the sampling line by your finger and check that the message ‘Sample line
blocked’ appears onto the monitor screen within 60 seconds.
?
25.
Detach the D-fend and check that the messages ‘Check D-fend’ appears onto the monitor
screen within 30 seconds.
?
Reattach the D-fend. Simulate at least 5 breaths by feeding calibration gas into the
sampling line. Check that the shown gas information is correct.
26.
Check that the monitor shows the message ‘Apnea’ within 30 seconds after you have
stopped feeding the gas.
?
27.
Turn the monitor off, disconnect and clean the module.
?
•
Fill in all necessary documents.
•
It is recommended to fill in the PM sticker, since the service check includes all the
Planned Maintenance actions. Attach it to a suitable place on the module box.