Compact Airway modules
Document No. 800 1009-1
23
3.2
Service check
These instructions include complete procedures for a service check. The service check is
recommended to be performed after any service repair. However, the service check procedures can
also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form,
which should be filled in when performing the
procedures.
The mark
?
in the instructions means that the check form should be signed after performing
the procedure.
The procedures are designed for monitors with software of level 00.
3.2.1
Recommended tools
Tool
Order No.
Notes
Screwdriver
Ambient pressure manometer
Flowmeter
Flow cassette 50/1.1
873812
Extra silicon tubing
Calibration gas and the regulator
755583 (gas)
755533
for M-CAiOVX/M-CAiOV/M-CAiO
Calibration gas and the regulator
755587 (gas)
755530
for M-COVX
Calibration gas and the regulator
755581 (gas)
755533
for M-COV/M-CO/M-C
Gas Interface Cable 2.5 m / 8 ft
884299
3.2.2
Recommended parts
Part
Order No.
Notes
Absorber
895933
D-fend
876446
D-fend+
881319
for M-COVX
Sampling line 3 m/10 ft
73319
anesthesia gas sampling line
Sampling line 2 m/7 ft
73318
for M-CAiOVX/M-COVX
D-lite / Pedi-lite
733950/73393
D-lite+
for condensing active humidification circuits