Datex-Ohmeda S/5 monitors
Document No. 800 1009-1
22
3
SERVICE PROCEDURES
3.1
General service information
Field service of the compact airway modules is limited to replacing faulty circuit boards or
mechanical parts. The circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full
type designation and a detailed fault description.
CAUTION
Only trained personnel with appropriate equipment should perform the tests and
repairs outlined in this section. Unauthorized service may void warranty of the unit.
CAUTION
The module electronics can only be repaired and calibrated at the factory.
3.1.1
OM measuring unit
CAUTION
Due to the complicated and sensitive mechanical construction any service inside
the O
2
measuring unit should not be attempted.
3.1.2
TPX measuring unit
CAUTION
The TPX photometer and its components are repaired/calibrated at the factory.
Attempts to repair/calibrate the unit elsewhere will adversely affect operation of the
unit. The information provided is for reference only.
3.1.3
OM, TPX, and PVX measuring unit
CAUTION
The OM, TPX, and PVX measuring units can be repaired only at the factory.
3.1.4
Serviceable or exchangeable parts
•
Absorber
•
D-fend
•
Nafion tubes
•
Fan filter
•
Fan
•
CPU board
•
CPU software
•
PVX Unit including PVX board
•
Pump
NOTE: After any component replacement see chapter