![Datex-Ohmeda CAPNOMAC ULTIMA Service Manual Download Page 34](http://html1.mh-extra.com/html/datex-ohmeda/capnomac-ultima/capnomac-ultima_service-manual_3771043034.webp)
CAPNOMAC ULTIMA
TM
SERVICE MANUAL
4-9
4.4.2
Principle of O
2
measurement
The differential oxygen measuring unit uses the paramagnetic principle
in a pneumatic bridge configuration. The signal picked up with a
differential pressure transducer is generated in a measuring cell with a
strong magnetic field that is switched on and off at a frequency of 110
Hz or 165 Hz. The output signal is a DC voltage proportional to the
O
2
concentration difference between the two gases to be measured.
January 15th, 1997/6
Summary of Contents for CAPNOMAC ULTIMA
Page 38: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 4 13...
Page 39: ...4 14 January 15th 1997 6...
Page 46: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 4 21...
Page 51: ......
Page 53: ......
Page 57: ......
Page 58: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 5 10...
Page 60: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 82: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 5 34 January 15th 1997 6...
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 108: ......
Page 109: ......
Page 111: ......
Page 117: ...ASX Preamplifier Board...
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 143: ......
Page 151: ......
Page 152: ......
Page 153: ......
Page 159: ......
Page 165: ......
Page 166: ......
Page 183: ...CAPNOMAC ULTIMA TM SERVICE MANUAL Other external parts January 15th 1997 6...
Page 185: ...CAPNOMAC ULTIMA TM SERVICE MANUAL 9 6 January 15th 1997 6...
Page 189: ...r 71 I I r a...
Page 190: ......
Page 211: ...DATEX CAPNOMAC ULTIMA SERVICE MANUAL 5 34 November 1st 1993 4...
Page 214: ......
Page 217: ......
Page 218: ......
Page 231: ...CAPNOMAC ULTIMA SERVICE MANUAL 12 13 January 15th 1997 6...