MATRIX 400™ CODE QUALITY VERIFIER SOLUTION
1
STEP 2 – CONNECT THE SYSTEM
To install the Matrix 400™ reader in a symbol verifier system configuration, you need the
hardware indicated in Figure 3 or Figure 4. In this layout the data is transmitted to the Host
on the main serial interface.
The RS232 auxiliary interface can be used for symbol verifier system configuration by
connecting an host computer running VisiSet™.
Refer to Matrix Series Ethernet Service Guide to connect the symbol verifier system to an
host computer by Ethernet TCP/IP interface (Matrix 400-x00-010 models only).
ISO/IEC 15415 AND ISO/IEC 15416 VERIFICATION
LTC-630
PG 6000
CAB-MS01
CBX
Main Interface
Host
Matrix 400™
BK-4990
LT-630
Code Surface
Figure 3 - Matrix 400™ in ISO/IEC 15415 and ISO/IEC 15416 Verifier Layout
1. Connect the Matrix 400™ to the CBX connection box.
2. Connect the LT-630 Four Bar Lighting System wires to the LTC-630 Power Supply
Controller through the adapter cable included in the LTC-630 package. Optionally, one of
the Matrix digital outputs can be used to switch the illuminator on/off at the LTC-630.
3. Position the “Coarse” and “Fine” regulations of the LTC-630 Power Supply Controller
respectively to “
6
” and “
9
” which are the preferred setting for symbol verification.
4. Connect the CBX to the PG600x power supply unit.
5. Connect the selected communication interface to the Host.
6. Connect the main power supply and switch on the system.
NOTE
For ISO/IEC 15415 and ISO/IEC 15416 verification, check the correct
orientation of the four LT-630 lighting bars (
45°
) before to perform the
verifier system calibration procedure described in
Step 7
.
4