Appendix A
84
Noise
shows the maximum noise measurement at 10 Hz with a gain of 1.
Figure 17: Maximum Noise Measurement During Typical Usage at 10 Hz with a Gain of 1
shows the typical RMS noise, in microvolts, of the DT9824 at sampling rates of
10 Samples/s, 100 Samples/s, 960 Samples/s, and 4800 Samples/s and gains of 1, 8, 16, and
32.
shows this data in graphical form.
Table 23: DT9824 System Noise
Gain
10 Samples/s
100 Samples/s
960 Samples/s
4800 Samples/s
1
3.6
μ
Vrms
12.0
μ
Vrms
25.0
μ
Vrms
63.0
μ
Vrms
8
0.7
μ
Vrms
2.0
μ
Vrms
4.2
μ
Vrms
11.0
μ
Vrms
16
0.6
μ
Vrms
1.8
μ
Vrms
2.8
μ
Vrms
7.5
μ
Vrms
32
0.4
μ
Vrms 1.3
μ
Vrms
2.3
μ
Vrms
6.5
μ
Vrms