Calibration
73
d.
Once the Running Average value has settled, adjust the gain by entering values in the
Gain edit box, or by clicking the up/down buttons until the Running Average value
matches the value you applied.
e.
Repeat steps 4a to 4d, selecting the next gain in the list, until the gain values are
calibrated for each gain setting.
5.
Repeat steps 2 to 4 for the remaining frequency until the offset and gain values are
calibrated for each frequency for the selected channel.
6.
Repeat steps 1 to 5 for the remaining A/D channels that you want to calibrate.
Note:
At any time, you can click
Restore Factory Settings
to reset the A/D calibration
values to their original factory settings. This process will undo any auto or manual calibration
settings.