DARAY X340 Operating & Installation Manual Download Page 36

 

Page 35 of 43 

5.3 Product Cleaning & Care Guidelines 

The X340 Range features BioProtect® antimicrobial coating on all painted surfaces and has a 
fully-enclosed head with an Ingress Protection (IP) rating of 54. 
These guidelines are provided to help you maintain your equipment properly and prevent any 
damage. Use of incompatible cleaning chemicals can adversely affect the finish and physically 
damage the metal and plastic parts of the DARAY product, invalidating its warranty.  

Please ensure these instructions are passed to the persons responsible for cleaning and they are 
followed before each use.  

Manufacturers recommended cleaning method: 

1.

 

Turn off the power to the light, and where possible, unplug the mains socket.  

2.

 

Remove noticeable contamination. Pay special attention to the light head and handle.  

3.

 

Ensure the equipment is surface-cleaned and disinfected using an appropriate, non-
abrasive cleaning agent (see recommended list below). 

4.

 

Dry all surfaces thoroughly with a clean cloth and check for damaged surfaces, 
discolouration and/or contamination.  
 

NOTE: Some DARAY products feature our BioProtect® antimicrobial coating on painted finishes. 
This is designed to assist with infection control and is not a replacement for cleaning.  

CAUTION: Our products are NOT waterproof. The equipment and accessories (included chargers 
and battery packs) MUST NOT at any point be saturated, to prevent water ingress. Use a damp 
cloth or saturated wipe if necessary. All electrical safety guidelines should be followed.  

Any damage caused as a result of not following these guidelines will void any product warranty.  

Below is a partial list of recommended surface agents: 

 

Cavicide™ products 

 

Isopropyl alcohol-based cleaners (<40% alcohol)  

 

Ethyl alcohol-based cleaners (<40% alcohol) 

 

Lysol® disinfectant  

 

Other non-bleach and non-abrasive disinfectants or cleaners 

 

DO NOT USE agents containing the following: 

 

Organic, mineral and oxidizing acids (min. pH value 8.5) 

 

Bases (max. acceptable pH value 8.5) 

 

Oxidation agents (e.g. Hydrogen Peroxide) 

 

Halogens (chlorine, iodine, bromide) 

 

Aromatic/halogenized hydrocarbons  

 

Bleach-based cleaners (e.g. Clorox™, Sterilox™) 

 

Abrasive cleaners (e.g. Comet Cleaner™) 

 

Acetone-based cleaners (e.g. nail polish remover, Goo-off™) 

 

MEK (Methyl Ethyl Ketone) 

If you are unsure whether you are using the correct cleaning fluids on your DARAY equipment, 
please feel free to get in touch with us for advice.

 

Summary of Contents for X340

Page 1: ...X340 LED Examination Light Operating Installation Manual QAM X340 0519 9...

Page 2: ...Page 1 of 43...

Page 3: ...ion 15 4 1 Wall mounted versions 15 4 1 1 Considerations 16 4 1 2 X340LW Placement and Installation 17 4 1 3 X340LE1 Placement and Installation 19 4 1 4 X340LE2 Placement and Installation 19 4 1 5 X34...

Page 4: ...llation 30 4 6 7 Light head and arm fixing 33 5 Maintenance 34 5 1 Maintenance schedule 34 5 2 Safety Precautions 34 5 3 Product Cleaning Care Guidelines 35 5 4 Detaching Parts 36 5 5 Fuses and their...

Page 5: ...e positioning of the light patch The light comprises 3 parts light head and articulated arm system power supply module and the mount of choice The rigid spring balanced arm allows the light patch to b...

Page 6: ...pported in this Manual X340LC X340 LED Ceiling Mounted Light X340LW X340 LED Wall Mounted Light X340LR X340 LED Rail Mounted Light X340LD X340 LED Desk Mounted Light X340LDS X340 LED Weighted Desk Sta...

Page 7: ...ing R9 88 Light Field Diameter D50 0 5m 90mm 0 75m 125mm 1 0m 175mm Light Field Diameter D10 0 5m 220mm 0 75m 255mm 1 0m 340mm Number of LED s 6 Power Consumption 14W 0 2A Average Working Life 50 000...

Page 8: ...V Plug top transformer 1 24V Plug top transformer 1 Desk Clamp Fixed Desk Mount Desk Stand 1 Wall bracket fixings 1 Light head and rigid arm system 1 Thumb Screw for Wall Mount 1 Microfibre cloth 1 Li...

Page 9: ...rigid arm system 1 User Manual Microfibre cloth 1 User Manual 1 2 2 Pre Start Checks Please check that the power cable is without visible damage harsh folds knots or cuts in the rubberised coating Wh...

Page 10: ...Page 9 of 43 3 Operation 3 1 Range of Motion The full range of motion between the X340 Range varies This is an image depicting the range of motion on a standard wall mount fitting X340LW...

Page 11: ...Page 10 of 43 X340LE4...

Page 12: ...e 11 of 43 This is an image depicting the range of motion on a standard ceiling mount fitting X340LC Downtube Rigid Arm Light head Tension Adjustment Downtube Extension Retainer Ring Ceiling Mount Cov...

Page 13: ...to power ON or OFF Press and hold the power button to adjust the intensity It takes approximately 5 seconds to switch between the highest and lowest settings Release the button when at a comfortable s...

Page 14: ...itself DO NOT confuse this with the tension adjustment knob on the left side see section 5 6 Page 35 Loosening this thumb screw will allow you to pivot the light head up or down The head has a total t...

Page 15: ...300mm see below image This extension is controlled by a thumb screw located around the back of the primary downtube Loosening the thumb screw will allow you to reposition the light to the required he...

Page 16: ...atible versions X340LW X340LE1 X340LE2 X340LE3 X340LE4 The wall bracket needs to be true on the vertical plane PIVOTW2 supplied with X340LW PIVOTW6 supplied with X340LE4 PIVOTW5 supplied with X340LE3...

Page 17: ...ical procedures the light will be used for If in doubt consult the end user or DARAY When installing the bracket take consideration of where you want the light head to sit when in its lowest position...

Page 18: ...ring in mind the final position of the light head required to project its light patch where it will be needed We recommend 850 1100mm from the floor to the underside of the bracket dependent on couch...

Page 19: ...ight on See page 12 Check arm and head for smooth and correct movement Your new DARAY light is now ready for use To gain full support of our warranty please fill in our warranty web form or contact DA...

Page 20: ...s version includes our PIVOTW4 spreader plate wall mounting bracket Suitable for fixing to all wall types including brickwork A separate power supply is included Mark on the wall the selected location...

Page 21: ...unting bracket Fit front bracket and rear plate to pre drilled panel Fit light spigot into bracket and pass cable through to the rear of panel Make electrical connection to transformer Fit pre drilled...

Page 22: ...s from the wall mounting box with the hex wrench provided Feed the mains supply via the cable entry hole on the back plate Connect the mains supply to the fused terminal block found on the back plate...

Page 23: ...il and fit the light to the rail The light can be moved along the rail by slackening the clamp slightly and not sufficiently to allow the light to become detached from the rail 4 3 Desk mounted versio...

Page 24: ...e upstand push down with a firm pressure until the upstand is fully inserted NOTE Insertion of upstand may be very tight and may require the use of a wooden or rubber mallet to tap the upstand downwar...

Page 25: ...e into the black power supply attached to the base upstand The light can now be tested by connecting the DC jacks between the power supply and the light and applying power Press the switch on the head...

Page 26: ...stand terminal Connect the plug top transformer to the 5521 IN OUT connector on the battery backup unit Insert the Plug top transformer into a 13amp wall socket Turn the battery back up unit on with t...

Page 27: ...und live circuitry should be qualified and following proper procedures Heavy or Difficult fittings Fitting the ceiling plate can be difficult and aspects of the fitting may require more than one perso...

Page 28: ...re and secured with chemical bonding The is our primary choice for mounting our lights to the ceiling structure appropriate rawl bolts and chemical bonding should be used to secure the fitting Bolted...

Page 29: ...Affixing the Ceiling plate The location of the light should be decided by the end user project manager or through schematical drawings Once the location has been decided it will be necessary to first...

Page 30: ...5mm twin and earth type DARAY recommends that the power supplied uses an earthed switched fused spur protected with a 3amp fuse The above wiring is for modern wiring applications post 2000 On completi...

Page 31: ...s of what length downtube would be required Height of Ceiling Plate from Floor Length of Downtube Required 2 4m 500 2 65m 750 2 9m 1000 3 4m 1500 3 9m 2000 The minimum length the downtube can be cut t...

Page 32: ...vent damage to the cable Before drilling the safety bolts hole please direct spigot on the bottom the downtube toward the centre of the examination area to allow the best range of motion over the exam...

Page 33: ...on the ceiling plate Slide the safety bolt through the clasps and the downtube being careful not to catch the wire Be sure to tighten the safety bolt and the nylon locknut It will be necessary to ens...

Page 34: ...ther The light can now be tested by applying power and pressing the power button on the back of the head Check the arm and head for smooth and correct movement Your new DARAY light is now ready for us...

Page 35: ...ge Remove arm clean spigot and re grease as necessary Check all electrical connections are sound and that there is no visible cable wear All fixings to be checked and be of sound construction 5 2 Safe...

Page 36: ...our BioProtect antimicrobial coating on painted finishes This is designed to assist with infection control and is not a replacement for cleaning CAUTION Our products are NOT waterproof The equipment...

Page 37: ...40 please contact DARAY as unauthorised work will invalidate the warranty 5 5 Fuses and their replacement For the ceiling mounted light the fuse is located on the ceiling plate on the terminal block c...

Page 38: ...arm appears to move by itself and does not settle in the required position This can be rectified by adjusting the spring tension on the arm Next to the light head on the left side of the arm systems...

Page 39: ...d the extension cable are fully connected For the mobile version ensure that the IEC is firmly pushed into the PSU Poor Light output Power Button Please check the variable intensity is not just set to...

Page 40: ...Amp ceiling plate fuse CS6144 For any SPARE PARTS or TECHNICAL ISSUES phone DARAY S Service Delivery Team on 44 0 800 878 9864 Opening times 9am 5pm Monday to Thursday excluding bank holidays 9am 2 30...

Page 41: ...red by any of the above We will always try to help but we cannot normally offer a refund For additional clarification please refer to our terms and conditions at www daray co uk terms In a small numbe...

Page 42: ...Daray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from...

Page 43: ...Page 42 of 43 Service Notes...

Page 44: ...Page 43 of 43...

Page 45: ...Page 44 of 43...

Page 46: ...Page 45 of 43...

Page 47: ...Page 46 of 43...

Page 48: ...DARAY Ltd Edison House Robian Way Swadlincote Derbyshire United Kingdom DE11 9DH 44 0 800 878 9864 44 0 333 321 0973 sales daray co uk www daray co uk...

Reviews: