![background image](http://html2.mh-extra.com/html/danfysik/81083852/81083852_manual_5125110007.webp)
SERVICE
5
MAGNET POWER SUPPLY SYSTEM 8500
Be aware when changing the Baud rates. Wrong setting may cause communication loss. Modifying
a baud rate with the HW switches will alter the baud rate right away, whereas modifying the baud
rate through the appropriate SW command, will first take place after a reset.
Reverting all set ups to the factory default can be accomplished as follows.
Short circuit ST60 (just below SW2). Press at the same time S2 continuously. All green
parameter LEDs will start flashing. After the fifth flash, the default parameters will be
restored. Please note that all parameters will be restored, also any selected addresses and the
ON pulse width. The last one may cause the power supply to be unable to be turned ON.
For switch settings, the following terminology will be used:
DANFYSIK A/S - DK 2630 TAASTRUP - DENMARK.
DOC NO P80303Sk