© Danfoss | DCS (MWA) | 2017.07
DKRCI.PI.SC0.F3.ML | 520H8392 | 5
Enter menu system
Enter QUICK SETUP
Unit change at
distance/level
readout:
m, cm, mm, in, ft
Change between:
Distance*
Level**
Output (%)***
Output (mA)****
4 – 20 mA output displayed as bar graph
and in percentage [%]
Measurement name (in this example, DISTANCE)
Device tag name
Measurement reading and unit
Device status (markers)
Marker 1, 2 and 3 (Error)
Hardware problem; the Signal Converter hardware is
defective. Contact Danfoss.
Marker 4 and 5 (Notification)
Depending on the level, the marker is ON or OFF.
Used for Danfoss service information only.
Keypad buttons
Flashing star indicating unit in operation.
* DISTANCE is a display option.
If the display is set to “DISTANCE” the displayed value will
be the distance from the Reference point to the top
surface of the liquid refrigerant (see fig. 5).
** LEVEL is display option.
If the display is set to “LEVEL” then the value displayed
will be:
PROBE LENGTH (entered in QUICK SETUP)
– DISTANCE (see fig. 5).
*** OUTPUT (%) is display option.
Will represent the level of refrigerant,in percent, scaled
(entered in QUICK SETUP) according to: SCALE 4 mA
(0%), SCALE 20 mA (100%) (see fig. 5).
**** OUTPUT I (mA) is display option.
Will represent the level of refrigerant,in 4-20
milliampere, scaled (entered in QUICK SETUP)
according to: SCALE 4 mA (4 mA), SCALE 20 mA
(20 mA) (see fig. 5).
*
AKS 4100
*
AKS 4100
12
12
DISTANCE
距离
进入Menue system (菜单系统)
进入 QUICK SETUP (快速设置)
距离/液位读数的单位设置:
m、cm、mm、in、ft
切换选项:
Distance(距离) *
Level(液位)液位**
Output(输出)(%)***
Output(输出)(mA)****
① 4 – 20 mA 输出(以柱状图和百分比[%]显示)
② 测量值名称(本例中为距离)
③ 设备名称
④ 测量读数和单位
⑤ 设备状态(标记)
标记1、2和3(错误)
硬件问题;信号转换器硬件有缺陷。请联系丹佛斯。
标记4和5(通知)
根据不同的液位,标记为开或关。只用于丹佛斯服
务信息。
⑥ 按键
⑦ 闪烁的星星,表示设备正在工作。
* “Distance(距离)”是一个显示选项。
若将显示设为“Distance”,则显示的值是参照点至
制冷剂液面顶部的距离(参考图 5)。
** “Level(液位)”是一个显示选项。
若将显示设为“Level”,则显示的值为:
探头长度(在快速设置中输入) – Level(参考图 5
)。
*** “Output(输出)(%)”是一个显示选项。
将以百分比形式显示制冷剂的液位,根据下列条件确
定标度(在快速设置中输入):
SCALE 4 mA(0%),SCALE 20 mA(100%)(参考图5)。
**** “Output(输出)(mA)”是一个显示
选项。
将以 4-20 毫安显示制冷剂的液位,根据下列条件确
定标度(在快速设置中输入):
SCALE 4 mA(4 mA),SCALE 20 mA
(20 mA)(参考图 5)。
en
zh