background image

User Guide 

SonoMeter 40

6 | BC389148615526en-010101

© Danfoss | Energy Meters | 2021.09

2.  Determination of 

measurement errors of 

the meter

2.1. Volume measuring 

errors determination test

The determination of volume measurement errors shall be carried out in the hydrodynamic test 

bench in the following order:

  1) The test mode is activated in accordance with section 1.1, 1.2, or 1.3 of this instruction.

  2) The volume measuring errors should be evaluated at control flow rates specified in EN1434-5. 

  3) The volume of water, passing through the meter can be read directly from the indicating device 

(with resolution 1 ml);

  - Via meter optical output, using the optical reading head according to 62056-21;

  - Or trough wired volume pulse 2nd output (for a complete meter with a connected pulse input /

output cable and for a meter activated in test mode according to p.1.2 of this instruction);

  4) Volume pulse values in test mode are presented in table 1p:

Table 1p

Permanent flow-rate qp of the heat meter, m³/h

Volume pulse value in test mode, litre/pulse

0,6 and 1,0

0,002

1,5

0,004

2,5

0,005

3,5 and 6

0,02

10; 15 and 25

0,05

40 and 60

0,2

2.2. Energy measurement 

errors determination test

The energy measurement error of a calculator with temperature sensors pair shall be evaluated by 

immersing the temperature sensors in a temperature regulated baths. The test shall be performed in 

the following order:

1) The test mode is activated in accordance with section 1.2 of this instruction;

2) The meter temperature sensors are immersed in thermostatic baths, which form the supply and 

return line temperature and temperature difference values specified in EN 1434-5. 

NOTE: energy measurement error determination may be performed separately for a calculator with 

a flow sensor. In this case, the temperature and temperature differences of the supply and return line 

specified in EN 1434-5 are simulated by connecting the reference resistors to the calculator terminals 

No.5;6;7;8.

3) Long press the button (for more than 5 seconds)  activates the simulation of the volume pulses (the 

meter display periodically shows "SF" with the nominal flow rate of the meter, m³/h):

4) After 2,5 min. the volume simulation is completed, the sign „SF“ turns off. To calculate the energy 

measurement error, the simulated volume and energy readings shall be visually read from the 

meter display;

5) The amount of volume or energy can be read through the wired pulse output (if it is equipped in 

the meter);

6) The amount of volume or energy can be read through the meter's optical interface output using an 

optical scan head that complies with EN 62056-21;

7) Energy pulse values in test mode are presented in table 2p:

Table 2p

Permanent flow-rate qp

of the heat meter, m³/h

Energy pulse value based on displayed energy units:

„kWh“, „MWh“

„GJ“

„Gcal“

0,6 and 1,0

0,1 Wh/pulse

0,5 kJ/ pulse

0,1 kcal/ pulse

1,5

0,2 Wh/ pulse

1 kJ/ pulse

0,2 kcal/ pulse

2,5

0,5 Wh/ pulse

2 kJ/ pulse

0,5 kcal/ pulse

3,5 and 6

1 Wh/ pulse

5 kJ/ pulse

1 kcal/ pulse

10; 15 and 25

2 Wh/ pulse

10 kJ/ pulse

2 kcal/ pulse

40 and 60

5 Wh/ pulse

20 kJ/ pulse

5 kcal/ pulse

0,6 and 1,0

10 Wh/ pulse

50 kJ/ pulse

10 kcal/ pulse

Summary of Contents for 40

Page 1: ...ergy Meters 2021 09 BC389148615526en 010101 1 SonoMeter 40 Ultrasonic energy meter for heating and cooling applications User Guide Ultrasonic energy meter SonoMeter 40 Test and calibration instruction...

Page 2: ...button 3 1 2 Activation of the test mode by short circuiting the contacts 4 1 3 Activation of test mode with software SonoMeter 40 Configurator 5 2 Determination of measurement errors of the meter 6...

Page 3: ...security password input window short press the button select digit in the first position after that long press the button and go to the next position after selecting the digit in the fourth position...

Page 4: ...TEST tESt on Wh Activated by pressing and holding the button To activate volume pulse output when energy pulse output is active TEST m3 tESt on Activated by pressing and holding the button 4 8 To dea...

Page 5: ...6 21 standard In this case optionally volume or energy pulses are generated via the optical interface of the meter ON TEST E Pulse mode intended for test TEST mode activation with energy pulse output...

Page 6: ...struction 2 The meter temperature sensors are immersed in thermostatic baths which form the supply and return line temperature and temperature difference values specified in EN 1434 5 NOTE energy meas...

Page 7: ...s activated in accordance with section 1 1 or 1 3 of this instruction NOTE the meter switches to the operating mode by itself 12 hours after activation the test mode 4 Meter calibration adjustment mod...

Page 8: ...s RESET Integrators and Loggers intended to reset the integrator and logger values to a zero RESET Battery time intended to reset the battery lifetime after replacement the new battery replacement dat...

Reviews: