This instruction is designed for the ultrasonic energy meter SonoMeter 40 for flow and energy 

verification and calibration.

General information

1. Activation of the test 

mode

1.2. Activation of the test 

mode by short-circuiting 

the contacts

The test mode can be activated in one of the following ways:

1.1. Activation of the test 

mode with the button

The test mode is activated by the meter’s control button according to the following procedure:

-  long press the button, on the meter’s LCD select page „INF“;

-  short press the button, select „TEST on Wh“ (when it is necessary to activate the energy pulse 

output via the optical interface) or „TEST On m³“ (when it is necessary to activate the volume pulse 

output via the optical interface);

-  long press the button, open the  4- digits security password input window:

-  short press the button, select digit in the first position, after that long press the button and go to 

the next position;

-  after selecting the digit in the fourth position, long press the button, the massage „PASS“ appears 

briefly (when the password entered correctly) and the meter switches to test mode – the sign 

TEST“ appears;

-  if the password was entered incorrectly, the message „FAIL“ appears briefly and the meter returns 

to the operating mode, and the procedure for turning on the test mode must be repeated initially;

-  the password value is fixed: 0001.
NOTE: when the test mode is activated by the button, the volume and energy accumulated in 

the test mode are added to the meter’s energy and volume readings in operating mode (after 

turning off the test mode).

By short-circuiting the contacts „SERVICE”, the SERVICE mode is activated, symbol „<->“ and sign 

TEST” are displayed on the LCD.

Remove the breakable partition „SERVICE” (1) on the back of the calculator or open the calculator box 

by removing the breakable partitions “LOCK” (2).

Summary of Contents for 40

Page 1: ...ergy Meters 2021 09 BC389148615526en 010101 1 SonoMeter 40 Ultrasonic energy meter for heating and cooling applications User Guide Ultrasonic energy meter SonoMeter 40 Test and calibration instruction...

Page 2: ...button 3 1 2 Activation of the test mode by short circuiting the contacts 4 1 3 Activation of test mode with software SonoMeter 40 Configurator 5 2 Determination of measurement errors of the meter 6...

Page 3: ...security password input window short press the button select digit in the first position after that long press the button and go to the next position after selecting the digit in the fourth position...

Page 4: ...TEST tESt on Wh Activated by pressing and holding the button To activate volume pulse output when energy pulse output is active TEST m3 tESt on Activated by pressing and holding the button 4 8 To dea...

Page 5: ...6 21 standard In this case optionally volume or energy pulses are generated via the optical interface of the meter ON TEST E Pulse mode intended for test TEST mode activation with energy pulse output...

Page 6: ...struction 2 The meter temperature sensors are immersed in thermostatic baths which form the supply and return line temperature and temperature difference values specified in EN 1434 5 NOTE energy meas...

Page 7: ...s activated in accordance with section 1 1 or 1 3 of this instruction NOTE the meter switches to the operating mode by itself 12 hours after activation the test mode 4 Meter calibration adjustment mod...

Page 8: ...s RESET Integrators and Loggers intended to reset the integrator and logger values to a zero RESET Battery time intended to reset the battery lifetime after replacement the new battery replacement dat...

Reviews: