background image

User Guide 

SonoMeter 40

BC389148615526en-010101 | 7

© Danfoss | Energy Meters | 2021.09

3. Turn off the test mode

The test mode can be turned off in one of the following ways:

-  long press the button selects page „INF“ on the meter’s LCD 

 short presses the button selects 

„tEST off“ on the LCD 

 long press the button and the test mode is turn off, there is no sign „TEST“ 

on the screen (when the test mode is activated in accordance with section 1.1 of this instruction);

-  by short-circuiting the contacts “SERVICE”, (when the test mode is activated in accordance with 

section 1.2 of this instruction);

-  via the optical interface, using the software “SonoMeter 40 Configurator” and optical head that 

complies with  EN 62056-21 standard (when the test mode is activated in accordance with section 

1.1 or 1.3 of this instruction);

NOTE: the meter switches to the operating mode by itself 12 hours after activation the test 

mode. 

4.  Meter calibration/

adjustment mode

Meter calibration/adjustment allows to adjust the meter measurement of volume single point 

characteristics.

It can be done by using the software “SonoMeter 40 Configurator” and optical scan head in 

accordance with EN 62056-21 standard.

1) The calibration/adjustment mode can be activated by removing protecting lid (2) of ADJ and 

shortcutting the pins.

2) The correction parameter for volume can be entered in the SonoMeter 40 Configurator fields 

“Err[%]”. The correction parameter is confirmed by clicking “Calculate”.  “Write configuration” stores 

the adjustments into the meter permanent memory. “Read configuration” is used to verify that the 

changes are stored.

NOTE: With the removed ADJ protective cover. The Manufacturer warranty is void!!!

Summary of Contents for 40

Page 1: ...ergy Meters 2021 09 BC389148615526en 010101 1 SonoMeter 40 Ultrasonic energy meter for heating and cooling applications User Guide Ultrasonic energy meter SonoMeter 40 Test and calibration instruction...

Page 2: ...button 3 1 2 Activation of the test mode by short circuiting the contacts 4 1 3 Activation of test mode with software SonoMeter 40 Configurator 5 2 Determination of measurement errors of the meter 6...

Page 3: ...security password input window short press the button select digit in the first position after that long press the button and go to the next position after selecting the digit in the fourth position...

Page 4: ...TEST tESt on Wh Activated by pressing and holding the button To activate volume pulse output when energy pulse output is active TEST m3 tESt on Activated by pressing and holding the button 4 8 To dea...

Page 5: ...6 21 standard In this case optionally volume or energy pulses are generated via the optical interface of the meter ON TEST E Pulse mode intended for test TEST mode activation with energy pulse output...

Page 6: ...struction 2 The meter temperature sensors are immersed in thermostatic baths which form the supply and return line temperature and temperature difference values specified in EN 1434 5 NOTE energy meas...

Page 7: ...s activated in accordance with section 1 1 or 1 3 of this instruction NOTE the meter switches to the operating mode by itself 12 hours after activation the test mode 4 Meter calibration adjustment mod...

Page 8: ...s RESET Integrators and Loggers intended to reset the integrator and logger values to a zero RESET Battery time intended to reset the battery lifetime after replacement the new battery replacement dat...

Reviews: