background image

User Guide 

SonoMeter 40

4 | BC389148615526en-010101

© Danfoss | Energy Meters | 2021.09

1.2. Activation of the test 

mode by short-circuiting 

the contacts 

(continuous)

Display readings in the test mode

ID

Parameter

Value

Notes

4.1

High- resolution energy

TEST

000000.00     Wh

Updated every second.

Indicated as “PULSE”, if the energy 

test pulse output is activated.

TEST

PULSE

4.2

High-resolution integrated volume

TEST

           m

3

00.000000

Updated every second.

Indicated as “PULSE m³”, if the 

volume test pulse output is 

activated.

TEST

           m

3

PULSE

4.3

Supply heat carrier temperature 

value

1                        TEST

0.0 °C

-

4.4

Return heat carrier temperature 

value

2                        TEST

0.0 °C

-

4.5

Temperature difference

1-2                    TEST

0.00 °C

-

4.6

High-resolution flow rate

TEST

           m

3

0.000

      

 INF

-

4.7

To activate energy pulses output 

(when volume pulse output is active)

TEST                                       

tESt on     Wh

Activated by pressing and holding 

the button

To activate volume pulse output  

(when energy pulse output is active)

   TEST    m

3

                     

tESt on  

       

Activated by pressing and holding 

the button

4.8

To deactivate the test mode 

TEST 

                     

tESt OFF        

Deactivated by pressing and holding 

the button

In this mode:

-  volume pulses are generated via the optical interface of the meter. The button can be used to 

toggle the energy  pulse output by selecting the menu  item "tEST on Wh”;

-  When the meter is supplied with a connected pulse input/output cable , the energy pulses are 

generated in the 1st pulse output and the volume pulses in the 2nd pulse output;

-  it is possible to simulate volume pulses for determination the energy measurement errors;

-  it is possible to change the parameters of the meter configuration. 
NOTE: when the test mode is activated by short-circuiting the contacts “SERVICE”, the volume 

and energy accumulated in the test mode are not added to the meter’s operating mode 

volume and energy readings.

Summary of Contents for 40

Page 1: ...ergy Meters 2021 09 BC389148615526en 010101 1 SonoMeter 40 Ultrasonic energy meter for heating and cooling applications User Guide Ultrasonic energy meter SonoMeter 40 Test and calibration instruction...

Page 2: ...button 3 1 2 Activation of the test mode by short circuiting the contacts 4 1 3 Activation of test mode with software SonoMeter 40 Configurator 5 2 Determination of measurement errors of the meter 6...

Page 3: ...security password input window short press the button select digit in the first position after that long press the button and go to the next position after selecting the digit in the fourth position...

Page 4: ...TEST tESt on Wh Activated by pressing and holding the button To activate volume pulse output when energy pulse output is active TEST m3 tESt on Activated by pressing and holding the button 4 8 To dea...

Page 5: ...6 21 standard In this case optionally volume or energy pulses are generated via the optical interface of the meter ON TEST E Pulse mode intended for test TEST mode activation with energy pulse output...

Page 6: ...struction 2 The meter temperature sensors are immersed in thermostatic baths which form the supply and return line temperature and temperature difference values specified in EN 1434 5 NOTE energy meas...

Page 7: ...s activated in accordance with section 1 1 or 1 3 of this instruction NOTE the meter switches to the operating mode by itself 12 hours after activation the test mode 4 Meter calibration adjustment mod...

Page 8: ...s RESET Integrators and Loggers intended to reset the integrator and logger values to a zero RESET Battery time intended to reset the battery lifetime after replacement the new battery replacement dat...