Danatronics, Corp.
20
Proper alignment of the target ball to the tip of the probe is critical for accurate readings.
Examples of good and poor ball alignment relative to the tip of the probe are shown
below and should occur only after the power up warm up time has been done and a
proper multi-point calibration performed
Note:
Proper alignment of the target ball to the tip of the probe is critical for accurate
readings.
Note:
The target ball size used to calibrate the MTG-99 must be used for all readings
until the gage is calibrated to another target ball size.
To begin the standard calibration, follow these steps:
1.
Connect the probe to the coiled cable and attach to the gage prior to turning on
the gage
2.
Power on the gage
3.
Wait 15 minutes for the gage and probe to warm up following the timer on the
screen.
4.
Press the
MENU/OK
key to access the
Main Menu
.
5.
Use the up and down arrow keys to select the
Calibration
option.
6.
Press the
MENU/OK
key.
7.
Use the up and down arrow keys to select the
Do Calibration
option.